OT PING METER
Report
- Report Number
- 2939301-2015-04533
- Event Type
- Malfunction
- Date Received
- February 6, 2015
- Report Date
- January 20, 2015
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K082590
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 ¿ (03/10/2015).THE PATIENT¿S METER HAS BEEN RETURNED ON 2/17/2015 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 2/27/2015 WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2015, THE REPORTER CONTACTED LIFESCAN USA ALLEGING THAT THE SUBJECT METER (ONETOUCH PING) READ INACCURATELY HIGH COMPARED TO ANOTHER DEVICE ON THREE OCCASIONS. THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF "299, 336 AND 350MG/DL" RESPECTIVELY WITH THE SUBJECT METER AND "111, 167 AND 248MG/DL" RESPECTIVELY ON ANOTHER DEVICE. ALL THREE COMPARISONS WERE PERFORMED SEPARATELY, BUT EACH METER/METER COMPARISON WAS PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR ACCURACY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.