RENEW RECEIVER KIT, 16-CHANNEL
Report
- Report Number
- 1627487-2015-12056
- Event Type
- Injury
- Date Received
- February 4, 2015
- Date of Event
- November 28, 2014
- Report Date
- April 6, 2015
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- K992946
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
RESULTS: THE REPORT FOR ¿INEFFECTIVE STIMULATION¿ WAS CONFIRMED. THE RENEW RECEIVER FAILED MANUAL TESTING FOR NO OUTPUT. THE ASIC¿S AND/OR OTHER INTERNAL COMPONENTS HAVE FAILED CAUSING THE NO OUTPUT CONDITION. THE RECEIVER WAS IMPLANTED FOR APPROXIMATELY 11 YEARS (3,931 DAYS). SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
PMA/510(K) NUMBER SHOULD BE K992946. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
(B)(4). SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
IT WAS REPORTED THE PATIENT IS UNABLE TO INCREASE STIMULATION WHICH WAS VERIFIED BY AN SJM REPRESENTATIVE. THE PATIENT HAS NOT USED THE SYSTEM IN OVER A YEAR. THE PATIENT'S IPG WAS EXPLANTED AND REPLACED. THE PATIENT REPORTEDLY RECEIVES EFFECTIVE STIMULATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 82430 | RENEW RECEIVER KIT, 16-CHANNEL | SCS IPG | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3416 | 26016A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Other | MODEL 3186(2), SSC LEAD |