FDA Adverse Event Injury Summary report: N

RENEW RECEIVER KIT, 16-CHANNEL

MDR report key: 4485028 · Received February 4, 2015

Report

Report Number
1627487-2015-12056
Event Type
Injury
Date Received
February 4, 2015
Date of Event
November 28, 2014
Report Date
April 6, 2015
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
GZB
PMA / PMN Number
K992946
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS: THE REPORT FOR ¿INEFFECTIVE STIMULATION¿ WAS CONFIRMED. THE RENEW RECEIVER FAILED MANUAL TESTING FOR NO OUTPUT. THE ASIC¿S AND/OR OTHER INTERNAL COMPONENTS HAVE FAILED CAUSING THE NO OUTPUT CONDITION. THE RECEIVER WAS IMPLANTED FOR APPROXIMATELY 11 YEARS (3,931 DAYS). SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Additional Manufacturer Narrative · 1

PMA/510(K) NUMBER SHOULD BE K992946. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Additional Manufacturer Narrative · 1

(B)(4). SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT IS UNABLE TO INCREASE STIMULATION WHICH WAS VERIFIED BY AN SJM REPRESENTATIVE. THE PATIENT HAS NOT USED THE SYSTEM IN OVER A YEAR. THE PATIENT'S IPG WAS EXPLANTED AND REPLACED. THE PATIENT REPORTEDLY RECEIVES EFFECTIVE STIMULATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
82430 RENEW RECEIVER KIT, 16-CHANNEL SCS IPG GZB ST. JUDE MEDICAL - NEUROMODULATION 3416 26016A

Patients

Seq Age Sex Outcome Treatment
1 75 YR Other MODEL 3186(2), SSC LEAD