3.5FR P.U.R. UMBIL CATH X10
Report
- Report Number
- 3009211636-2015-00047
- Event Type
- Malfunction
- Date Received
- February 4, 2015
- Date of Event
- January 14, 2015
- Report Date
- January 21, 2015
- Manufacturer
- COVIDIEN
- Product Code
- FOS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
SUBMIT DATE: 2/04/2015. AN INVESTIGATION IS CURRENTLY UNDERWAY. UPON COMPLETION, THE RESULTS WILL BE FORWARDED.
THE LOT NUMBER WAS PROVIDED AND THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED INDICATING THAT THERE WAS NO QUALITY ISSUES ASSOCIATED WITH THIS DEFECT MODE. THERE ARE NO NON- CONFORMANCES RELATED TO THE REPORTED ISSUE. THE PRODUCT SAMPLE WAS NOT RETURNED AND NO PHOTOS WERE RECEIVED. BECAUSE A SAMPLE WAS NOT RETURNED, WE WERE UNABLE TO PERFORM A THOROUGH FOLLOW UP INVESTIGATION TO INCLUDE FUNCTIONAL AND VISUAL EVALUATIONS TO CONFIRM THE DEFECT AND ROOT CAUSE ANALYSIS. THIS COMPLAINT WILL BE REOPENED AND UPDATED ACCORDINGLY IF THE PRODUCT SAMPLE IS RETURNED. THE CATHETER WAS REPORTED TO HAVE FUNCTIONED AS INTENDED FOR APPROXIMATELY 2 WEEKS, THEREFORE IT IS CONSIDERED THAT MORE LIKELY, THE DEVICE WAS DAMAGED DURING THAT TIME. WITH THE AVAILABLE INFORMATION, THE MOST PROBABLE ROOT CAUSE COULD BE CONSIDERED AS MISUSE (LEAK COULD BE CAUSED DUE TO OVER BENDING OR EXCESSIVE FORCE). HOWEVER WITHOUT THE SAMPLE, THIS CANNOT BE CONFIRMED. THE LOT INVOLVED WITHIN THIS COMPLAINT WAS MANUFACTURED ON 5/3/2013 PRIOR TO IMPLEMENTATION OF SPECIAL 510K#K130725. UNDER SPECIAL 510K#K130725, COVIDIEN EXPANDED WARNINGS WITHIN THE INSTRUCTIONS FOR USE MANUAL OF THIS PRODUCT TO HIGHLIGHT THE CONCERN OF UVC CATHETER DAMAGE IN THE HUB AREA ASSOCIATED WITH THE USE OF ALCOHOL BASED DISINFECTANTS. THE INSTRUCTIONS FOR USE WERE REPLACED IN ALL STOCK, AND HOSPITALS WERE SENT A LETTER HIGHLIGHTING THIS ISSUE. IT MUST BE NOTED THAT THE INSTRUCTIONS FOR USE (IFU) STATES: DO NOT USE INSTRUMENTS WITH SHARP OR ROUGH EDGES DIRECTLY ON THE CATHETER SINCE EVEN A MINOR CUT COULD TEAR OR BREAK THE CATHETER AND CONTINUES, DO NOT PINCH OR BEND THE CATHETER BACK TO TEMPORARILY OCCLUDE THE CATHETER. THIS CAUSES INCREASED STRESS ON THE CATHETER WHICH CAN LEAD TO A LEAK OR BREAK. DO NOT USE ALCOHOL, ACETONE, OR ALCOHOL CONTAINING ANTISEPTICS DIRECTLY ON THE CATHETER. CAREFULLY CHECK ANTISEPTIC SOLUTIONS FOR ALCOHOL OR ACETONE. THESE SUBSTANCES MAY CAUSE...
IRREVERSIBLE DAMAGE TO THE POLYURETHANE WHICH CAN LEAD TO A LEAK OR BREAK. ENSURE GLOVES OR OTHER SURFACES WHICH HAVE ALCOHOL ON THEM ARE COMPLETELY DRY BEFORE TOUCHING OR MANIPULATING THE CATHETER. ALSO UNDER THE SPECIAL 510K#K130725, A DESIGN CHANGE WAS IMPLEMENTED TO INCREASE THE LENGTH OF THE STRAIN RELIEF ON THE LUER HUB OF THE SINGLE LUMEN UVC CATHETERS. THIS CHANGE IS EXPECTED TO REDUCE THE STRESS AND LIKELIHOOD OF KINKING IN THE AREA MOST LIKELY TO BE EXPOSED TO ALCOHOL. THIS CHANGE WAS EXECUTED IN MAY 2014. ADDITIONALLY IN-PROCESS CONTROLS (SUCH AS PERSONNEL TRAINING, INCOMING QUALITY ACCEPTANCE TESTING FOR RAW MATERIAL, 100% IN PROCESS VISUAL INSPECTION AND VISUAL ACCEPTANCE SAMPLING) ARE IN PLACE TO PREVENT NONCONFORMING PRODUCT FROM LEAVING THE MANUFACTURING OPERATIONS. MANUFACTURING PERFORMS 100% LEAK TESTING AT THE FINAL STAGE OF PRODUCTION, WHICH WOULD IDENTIFY THIS ISSUE IN THE CATHETER ASSEMBLY. THIS COMPLAINT WILL BE USED FOR TRACKING AND TRENDING PURPOSES.
IT WAS REPORTED TO COVIDIEN ON 01/21/2015 THAT A CUSTOMER HAD AN ISSUE WITH A SINGLE LUMEN UMBILICAL VESSEL CATHETER (UVC). THE CUSTOMER STATES THAT A FISSURE(CRACK) WAS FOUND ON THE CATHETER AND A NEW UNIT WAS INSERTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 82533 | 3.5FR P.U.R. UMBIL CATH X10 | UMBILICAL VESSEL CATHETER | FOS | COVIDIEN | 8888160333 | 312302X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 DA |