FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 4474365 · Received July 21, 2014

Report

Report Number
1314492-2014-45018
Event Type
Malfunction
Date Received
July 21, 2014
Date of Event
March 1, 2014
Report Date
July 7, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

MFR REF NO: (B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT 152305. THE SYSTEM ERROR 322 ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADD'L INFO IS RECEIVED, A F/U WILL BE SENT. THE MOTOR SCREW, UPPER AUXILIARY ASSEMBLY AND LOWER AUXILIARY ASSEMBLY WERE REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE LOWER LINK SWITCH DOES NOT ACTIVATE. BAXTER HAS INITIATED A CAPA TO FURTHER INVESTIGATE SYSTEM ERROR 322 ALARMS.

Description of Event or Problem · 1

DURING THE EVAL OF A RETURNED SPECTRUM INFUSION PUMP, 5 OCCURRENCES OF "SYSTEM ERROR 322" ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNK SINCE THESE ITEMS WERE FOUND DURING EVAL OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
426743 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1