MCKESSON ANESTHESIA CARE
Report
- Report Number
- 3009662297-2015-00001
- Event Type
- Malfunction
- Date Received
- January 23, 2015
- Date of Event
- January 22, 2015
- Report Date
- January 27, 2015
- Manufacturer
- MCKESSON TECHNOLOGIES INC.
- Product Code
- BSZ
- PMA / PMN Number
- K113633
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
UPON INVESTIGATION, IT WAS DETERMINED THAT WHEN THE ANESTHESIA CARE RECORD (ACR) APPLICATION IS INSTALLED ON A WORKSTATION RUNNING (B)(4), THE ACR APPLICATION WILL FAIL TO RETAIN CASE DATA IF THE USER HOVERS HIS/HER MOUSE OVER THE NOTES (COMMENTS) PANE, THE MESSAGES PANE, OR THE FLUID INTAKE/OUTPUT SCREEN AND A TOOL TIP IS DISPLAYED. AS A RESULT, DURING THIS PERIOD OF INACTIVITY, CASE DATA FROM THE PHYSIOLOGICAL MONITORS ARE NOT RECORDED AND CASE DATA UPDATES RECEIVED FROM OTHER SYSTEMS (VIA INTERFACES AND INTEGRATION) ARE NOT RETAINED ON THE WORKSTATION. WHEN THE USER¿S MOUSE IS RE-POSITIONED, THE APPLICATION WILL RETURN TO ITS ACTIVE STATE AND BEGIN TO RETAIN CASE DATA. THE REPORTING HOSP SITE WAS NOTIFIED OF THE RESULTS OF THE INVESTIGATION AND THE POTENTIAL OF THE ACR SCREEN TO BECOME UNRESPONSIVE AND DATA TO NOT RECORD. TYPICALLY THE ACR APPLICATION IS USED AS TOUCH-SCREEN APPLICATION WHERE THE MOUSE IS RARELY UTILIZED. THIS EVENT WILL NOT OCCUR ON A WORKSTATION WHEN THE USER IS USING A TOUCH-SCREEN MONITOR WITHOUT A MOUSE. THERE WAS NO INJURY ON THE ACTUAL PT AND/OR NO DELAY IN PT CARE.
IT WAS REPORTED BY ONE CUSTOMER SITE OF AN ALLEGED LOSS OF DATA DISPLAY BETWEEN OUR COMPANY¿S SOFTWARE APPLICATION AND ANOTHER VENDOR¿S PHYSIOLOGICAL DEVICE. THERE WERE NO ADVERSE EVENTS OF PT INJURIES RESULTANT FROM THE REPORTED ALLEGED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 54041 | MCKESSON ANESTHESIA CARE | MCKESSON ANESTHESIA CARE | BSZ | MCKESSON TECHNOLOGIES INC. | MAX 15.X VERSIONS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |