FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 4463832 · Received July 2, 2014

Report

Report Number
1314492-2014-41020
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
June 10, 2014
Report Date
June 10, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Removal / Correction Number
1314492-080712-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED AND REPRODUCED THE REPORTED SYMPTOM DYSFUNCTIONAL KEYPAD. THE EVALUATION FOUND THAT ALL OF THE KEYS HAD AN INTERMITTENT RESPONSE. THE FAILED KEYPAD WAS REPLACED. BAXTER HAS INITIATED A CAPA TO FURTHER INVESTIGATE THE ISSUE.

Description of Event or Problem · 1

I TWAS REPORTED THAT A SPECTRUM INFUSION PUMP HAD A DYSFUNCTIONAL KEYPAD. TI WAS ALSO REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
388915 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1