FDA Adverse Event Malfunction Summary report: N

CLEARSIGN? AMPLIFIER

MDR report key: 4461056 · Received January 28, 2015

Report

Report Number
2134265-2015-00590
Event Type
Malfunction
Date Received
January 28, 2015
Date of Event
December 16, 2014
Report Date
December 30, 2014
Manufacturer
BOSTON SCIENTIFIC - LOWELL, MA
Product Code
DRQ
PMA / PMN Number
K050006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE RESULTS OF THE TESTING INDICATED THAT THE RETURNED CARTO CABLE AS WELL AS THE ASSOCIATED CLEARSIGN AMPLIFIER PASSED INCOMING INSPECTION TESTS, INCLUDING NOISE AND LEAKAGE TESTING. THE CABLE SHIELDING CONNECTION WAS VERIFIED BY MEASURING THE CAPACITANCE AND IMPEDANCE AND RESULTS WERE SIMILAR TO THE CONTROL CABLE. THE POSSIBILITY OF INTERACTION EFFECTS BETWEEN THE TWO COMPONENTS AS A CAUSE WAS ELIMINATED BY TESTING THE RETURNED COMPLAINT AMPLIFIER AND CABLE TOGETHER AND FOUND NO ISSUES. BOTH DEVICES PERFORMED AS EXPECTED. THE BATCH NUMBER IS UNKNOWN AND THE MANUFACTURING RECORDS FOR THE COMPLAINT DEVICE COULD NOT BE REVIEWED. THE MOST PROBABLE CAUSE OF THE REPORTED DIFFICULTIES MAY BE DUE INTERACTION WITH ANOTHER DEVICE. (B)(4).

Description of Event or Problem · 1

SAME CASE AS MDR 2134265-2015-00348. IT WAS REPORTED THAT UNINTENDED RF ENERGY WAS DELIVERED TO THE PATIENT. DURING AN ABLATION PROCEDURE, 10HZ NOISE WAS SEEN ON THE ABLATION CHANNEL WHEN NO RF ENERGY WAS APPLIED. SOMEHOW THERE WAS PACING CAPTURED IN THE ATRIUM FOR THE ABLATION CATHETER (10HZ), INDICATING THE ABLATION CATHETER WAS SENDING ENERGY INTO THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT IS IN GOOD CONDITION. NO ADDITIONAL TREATMENT WAS REQUIRED.

Description of Event or Problem · 1

SAME CASE AS MDR 2134265-2015-00348. IT WAS REPORTED THAT UNINTENDED RF ENERGY WAS DELIVERED TO THE PATIENT. DURING AN ABLATION PROCEDURE, 10HZ NOISE WAS SEEN ON THE ABLATION CHANNEL WHEN NO RF ENERGY WAS APPLIED. SOMEHOW THERE WAS PACING CAPTURED IN THE ATRIUM FOR THE ABLATION CATHETER (10HZ), INDICATING THE ABLATION CATHETER WAS SENDING ENERGY INTO THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT IS IN GOOD CONDITION. NO ADDITIONAL TREATMENT WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
64849 CLEARSIGN? AMPLIFIER AMPLIFIER AND SIGNAL CONDITIONER, TRANSDUCER SIGNAL DRQ BOSTON SCIENTIFIC - LOWELL, MA H30120019680 EPVI0094

Patients

Seq Age Sex Outcome Treatment
1 CLEARSIGN AMPLIFIER| SMART TOUCH CATHETER| STOCKERT GENERATOR, BIOSENSE WEBSTER