FDA Adverse Event Injury Summary report: N

ALTERG ANTI-GRAVITY TREADMILL

MDR report key: 4459680 · Received January 26, 2015

Report

Report Number
3007033433-2015-00001
Event Type
Injury
Date Received
January 26, 2015
Date of Event
January 16, 2015
Report Date
January 22, 2015
Manufacturer
ALTERG, INC
Product Code
IOL
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALTERG INC. WILL ADDRESS THE ISSUE BY NOTIFYING ALL CUSTOMERS ON THE IMPORTANCE OF ENSURING PATIENT STABILITY BEFORE ENTRY/EXIT OF ALTERG EQUIPMENT TO PROTECT PATIENTS FROM FOLLOWING TRANSITIONS.

Description of Event or Problem · 1

A PATIENT WAS INJURED EXITING THE F320 PRODUCT AFTER A WORKOUT SESSION. ACCORDING TO THE DIRECTOR OF REHAB AT THE FACILITY, THE PATIENT HAS JUST ENDED HIS SESSION AND WAS HAVING HIS VITALS TAKEN. WHILE EXITING THE TREADMILL, THE COCKPIT WAS UNLOCKED; THE PATIENT WAS NOT VERTICALLY STABLE AND PUT WEIGHT ON HIS HEELS. THIS CAUSED THE TREADMILL BELT TO MOVE FORWARD, CAUSING HIS FEET TO SLIDE OUT FROM UNDER HIM. THE PATIENT COULD NOT STOP THE MOMENTUM, NOR COULD THE PT AIDING THE PATIENT. THE PATIENT SAT ON THE REAR OF THE COCKPIT, WHICH ADDED FURTHER WEIGHT AND MOMENTUM TO THE FALL, HIS LEGS FELL OUT IN FRONT OF HIM, AND WERE PINNED UNDER THE FRONT OF THE COCKPIT. NOTE THE PATIENT HAD BEEN IN THE EQUIPMENT ONCE BEFORE. HE WAS CONSIDERED A MINIMUM-ASSIST PATIENT, AND WAS WEARING A GAIT BELT. THE PT WAS TO THE SIDE OF THE EQUIPMENT, AND WAS UNABLE TO ASSIST THE PATIENT TO STAND-UP VERTICALLY ONCE THE PROCESS BEGAN, NOR TO STOP THE PROGRESS OF THE BELT SLIPPAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
58832 ALTERG ANTI-GRAVITY TREADMILL IOL, TREADMILL POWERED IOL ALTERG, INC F320

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention