FDA Adverse Event
Other
Summary report: N
WASHER STERILIZER
MDR report key: 44590
·
Received October 7, 1996
Report
- Report Number
- 2515984-1996-90001
- Event Type
- Other
- Date Received
- October 7, 1996
- Date of Event
- September 30, 1996
- Report Date
- October 4, 1996
- Manufacturer
- AMERICAN STERILIZER CO.
- Product Code
- FLE
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
SOAP EJECTOR WAS CLOGGED. INSTEAD OF WATER FLOWING THROUGH EJECTOR SIPHONING DETERGENT FROM CONTAINER, THE WATER FLOWED INTO THE SOAP CONTAINER AND OVERFLOWED SPILLING SOAPY WATER ONTO THE FLOOR. OPERATOR SLIPPED ON FLOOR BRUISING RIGHT ELBOW AND HIP. ALSO HIT HEAD ON FLOOR. THIS WAS A MINOR INJURY, BUT COULD HAVE BEEN A SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WASHER STERILIZER | RELIANCE 3400 WASHER STERILIZER | FLE | AMERICAN STERILIZER CO. | 3400 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |