ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2015-00059
- Event Type
- Malfunction
- Date Received
- January 19, 2015
- Date of Event
- August 22, 2014
- Report Date
- September 11, 2014
- Manufacturer
- CAREFUSION CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K950419
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). FIVE MDRS (MFR #S 2016493-2015-00056, 57, 58, 59, 60) ARE BEING SUBMITTED FOR THIS EVENT, ONE FOR EACH OF THE SUSPECT DEVICES INVOLVED. THE REPORTED OCCURRENCE OF A COMMUNICATION ERROR WAS CONFIRMED AND DUPLICATED ON THE RETURNED INFUSION SYSTEM. IN ADDITION, LOG ANALYSIS IDENTIFIED THREE OCCURRENCES OF CHANNEL DISCONNECTS WITH THE SYSTEM DURING USE ON (B)(6) 2014. IN THE FIRST OCCURRENCE THE TWO PUMP MODULES ATTACHED TO THE RIGHT SIDE OF THE PCU (CHANNELS C AND D) EXPERIENCED A POWER INTERRUPTION THAT STOPPED THE BASIC INFUSION RUNNING ON CHANNEL C (LVP SERIAL NUMBER (B)(4)) AND THE NITROGLYCERIN INFUSION RUNNING ON CHANNEL D (LVP SERIAL NUMBER (B)(4)). THE INFUSIONS WERE RESTORED AND RESTARTED BY THE USER. THE SECOND INCIDENT OCCURRED APPROXIMATELY 5 1/2 HOURS LATER IN CHANNEL A (LVP SERIAL NUMBER (B)(4)), INFUSING EPINEPHRINE AND ATTACHED TO THE LEFT OF CHANNEL B (LVP SERIAL NUMBER (B)(4)), WHICH WAS POSITIONED DIRECTLY TO THE LEFT OF THE PC UNIT. AT THAT TIME, CHANNEL A, EXPERIENCED A COMMUNICATION ERROR, WITH THE MODULE CONTINUING TO INFUSE AS DESIGNED. THE LOG RECORDED THE MODULE AS HAVING BEEN RE-ATTACHED, THE PROGRAMMING RESTORED, AND THE INFUSION RESTARTED. THE THIRD INCIDENT OCCURRED A FEW MINUTES LATER; A CHANNEL DISCONNECTED ALARM WAS RECORDED WHEN CHANNEL B, INFUSING NOREPINEPHRINE, EXPERIENCED A POWER INTERRUPTION AND CHANNELS A, C AND D EXPERIENCED A COMMUNICATION ERROR. THE USER ATTEMPTED TO RESTART THE NOREPINEPHRINE AFTER CLEARING THE ALARM STATE BUT RECEIVED ANOTHER CHANNEL DISCONNECTED ALARM AND SHUT DOWN THE ENTIRE SYSTEM. TESTING FOUND THE PRIMARY SOURCE FOR THE INCIDENTS THAT OCCURRED IN THE CHANNELS POSITIONED ON THE RIGHT SIDE OF THE PCU TO BE CAUSED BY THE RIGHT IUI CONNECTOR OF THE PCU. THE SOURCE FOR THE INCIDENTS THAT OCCURRED IN THE CHANNELS POSITIONED ON THE LEFT SIDE OF THE PCU WAS FOUND TO BE CAUSED BY BOTH THE LEFT AND THE RIGHT IUI CONNECTORS OF PUMP MODULE SN (B)(4) (CHANNEL B). WHEN THE NOTED IUI CONNECTORS WERE TEMPORARILY REPLACED, THE SYSTEM WAS FOUND TO HAVE NO REOCCURRENCE OF ANY CONTINUITY-RELATED ISSUES AT THE IUI CONNECTIONS, EVEN UNDER AGGRESSIVE PHYSICAL MANIPULATION OF THE SYSTEM. INSPECTION OF THE SOURCE IUI CONNECTORS FOUND THE PRESENCE OF THE CONTAMINATION/CORROSION ON PINS WHICH IS AN INDICATION THE PINS HAVE COME INTO CONTACT WITH A FLUID OTHER THAN (B)(4) ISOPROPYL ALCOHOL. THE PROXIMATE CAUSE FOR THE CUSTOMER'S REPORTED EXPERIENCE IS ATTRIBUTED TO CONTAMINATION/CORROSION ON THE PINS OF THE IUI CONNECTORS ON THE PC UNIT AND MULTIPLE ATTACHED MODULES. THE ROOT CAUSE FOR THE IUI CONNECTORS AS-RECEIVED IS AN INDICATION OF POSSIBLE IMPROVEMENT NEEDED IN THE CLEANING AND PM PRACTICES BEING PERFORMED.
THE CUSTOMER REPORTED THAT THEY HAD A PCU COMMUNICATION ERROR; THE SYSTEM CONSISTED OF A PCU WITH FOUR ATTACHED MODULES INITIALLY INFUSING NOREPINEPHRINE, EPINEPHRINE, AND NITROGLYCERINE. REPORTEDLY, WHEN THE COMM ERROR OCCURRED "ALL FUNCTIONS THEN FROZE AND THE PUMPS SHUT DOWN". THE PATIENT WAS ON AN INTRA-AORTIC BALLOON PUMP, AND WAS UNSTABLE PRIOR TO THE EVENT. A NEW PUMP AND MODULES WERE PROGRAMMED AND PUT INTO USE WITHIN FIVE MINUTES OF THE EVENT. NO PATIENT HARM OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 43981 | ALARIS PUMP MODULE | PUMP, INFUSION | FRN | CAREFUSION CORPORATION | 8100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | ALARIS PC UNIT, SN (B)(4)| ALARIS PUMP MODULE TUBINGS, MODEL/LOT UNKNOWN| ALARIS PUMP MODULE, SNS (B)(4)| ALARIS PUMP MODULE, SNS (B)(4)| ALARIS PUMP MODULE, SNS (B)(4) |