2.5MM DRILL BIT/QC/GOLD/180MM

MDR report key: 4447622 · Received January 23, 2015

Report

Report Number
3003506883-2015-10028
Event Type
Malfunction
Date Received
January 23, 2015
Date of Event
January 12, 2015
Report Date
January 12, 2015
Manufacturer
SYNTHES ELMIRA
Product Code
HWE
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: ADDITIONAL PRODUCT CODE: GFF, GFA, HSZ. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO COMPLAINT RELATED ANOMALIES. THE DEVICE HISTORY RECORD SHOWS LOT# 7080904 OF 2.5MM DRILL BIT/QC/GOLD/180MM WAS PROCESSED THROUGH THE NORMAL MANUFACTURING AND INSPECTION OPERATIONS WITH NO REWORK OR NONCONFORMITIES NOTED. THIS LOT MET ALL DIMENSIONAL AND VISUAL CRITERIA AT THE TIME OF MANUFACTURE WITH NO ISSUES DOCUMENTED DURING THE MANUFACTURE THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A DRILL BIT BROKE ONE CENTIMETER FROM THE DISTAL TIP. THE BROKEN PIECE COULD NOT BE RETRIEVED. THIS CASE WAS FOR THE DISTAL HUMERUS. PIECE WAS NOT IN THE JOINT. THERE WAS NO DELAY REPORTED. THIS REPORT IS 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56001 2.5MM DRILL BIT/QC/GOLD/180MM INSTR,SURGICAL,ORTHOPEDIC,AC-POWERED MOTOR/ACCESS & ATTACH HWE SYNTHES ELMIRA 7080904