DA VINCI SI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2015-00147
- Event Type
- Malfunction
- Date Received
- January 23, 2015
- Date of Event
- January 5, 2015
- Report Date
- January 15, 2015
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081137
- Removal / Correction Number
- 2955842-020312-001C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- OTHER
Narratives
THE PSM ARM WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THE PSM FAILED THE WEIGHTED BRAKE DROP TEST. INTUITIVE SURGICAL, INC. (ISI) HAS CONDUCTED A DEVICE HISTORY RECORD (DHR) REVIEW FOR THIS PSM AND DID NOT FIND ANY NON-CONFORMANCES THAT WOULD AFFECT ANY MATERIAL OF THE FINAL PRODUCT AND/OR THE QUALITY OR PERFORMANCE OF THE PSM USED DURING THIS REPORTED EVENT. THIS COMPLAINT IS BEING REPORTED DUE TO THE PSM ARM FAILING THE WEIGHTED BRAKE DROP TEST. ALTHOUGH THERE WAS NO PATIENT INVOLVEMENT, IT COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE MALFUNCTION WERE TO RECUR.
DURING THE PREVENTATIVE MAINTENANCE OF A DA VINCI SI SURGICAL SYSTEM, IT WAS OBSERVED THAT ONE OF THE PATIENT SIDE MANIPULATORS (PSM) FAILED THE VISCOUS FRICTION TEST ON AXIS 1. THE PSM IS AN INSTRUMENT ARM WHICH IS LOCATED ON THE PATIENT SIDE CART THAT PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENT. THE PSM WAS REPLACED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR ADVERSE OUTCOMES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 54736 | DA VINCI SI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS3000 A6.0P8 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |