FDA Adverse Event Malfunction Summary report: N

DA VINCI SI SURGICAL SYSTEM

MDR report key: 4446993 · Received January 23, 2015

Report

Report Number
2955842-2015-00147
Event Type
Malfunction
Date Received
January 23, 2015
Date of Event
January 5, 2015
Report Date
January 15, 2015
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K081137
Removal / Correction Number
2955842-020312-001C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PSM ARM WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THE PSM FAILED THE WEIGHTED BRAKE DROP TEST. INTUITIVE SURGICAL, INC. (ISI) HAS CONDUCTED A DEVICE HISTORY RECORD (DHR) REVIEW FOR THIS PSM AND DID NOT FIND ANY NON-CONFORMANCES THAT WOULD AFFECT ANY MATERIAL OF THE FINAL PRODUCT AND/OR THE QUALITY OR PERFORMANCE OF THE PSM USED DURING THIS REPORTED EVENT. THIS COMPLAINT IS BEING REPORTED DUE TO THE PSM ARM FAILING THE WEIGHTED BRAKE DROP TEST. ALTHOUGH THERE WAS NO PATIENT INVOLVEMENT, IT COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

DURING THE PREVENTATIVE MAINTENANCE OF A DA VINCI SI SURGICAL SYSTEM, IT WAS OBSERVED THAT ONE OF THE PATIENT SIDE MANIPULATORS (PSM) FAILED THE VISCOUS FRICTION TEST ON AXIS 1. THE PSM IS AN INSTRUMENT ARM WHICH IS LOCATED ON THE PATIENT SIDE CART THAT PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENT. THE PSM WAS REPLACED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR ADVERSE OUTCOMES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
54736 DA VINCI SI SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL,INC. IS3000 A6.0P8

Patients

Seq Age Sex Outcome Treatment
1