FDA Adverse Event
Death
Summary report: N
BIVONA TTS
MDR report key: 444471
·
Received February 10, 2003
Report
- Report Number
- 444471
- Event Type
- Death
- Date Received
- February 10, 2003
- Date of Event
- November 26, 2002
- Report Date
- February 10, 2003
- Manufacturer
- BIVONA MEDICAL TECHNOLOGIES
- Product Code
- BTO
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- FL, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
THE PT HAD THE BIVONA TRACH PLACED EMERGENTLY IN 2002 FOR RESPIRATORY DISTRESS. 5 DAYS LATER PT DEVELOPED DECREASED OXYGEN SATURATION, STAFF DISCOVERED TRACH WAS DISPLACED. RESUSCIATION EFFORTS FAILED. INVESTIGATION REVEALED THE SLIDING LOCKABLE NECK FLANGE FAILED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BIVONA TTS | ADJUSTABLE NECK FLANGE HYPERFLEX TRACH TUBE | BTO | BIVONA MEDICAL TECHNOLOGIES | CE 0473 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR | Death |