FDA Adverse Event Malfunction Summary report: N

ACTIVA

MDR report key: 4438649 · Received January 21, 2015

Report

Report Number
3004209178-2015-01036
Event Type
Malfunction
Date Received
January 21, 2015
Report Date
December 30, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
MHY
PMA / PMN Number
P960009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT: PRODUCT ID 37642, SERIAL# (B)(4), PRODUCT TYPE PROGRAMMER, PATIENT. PRODUCT ID 37085-60, SERIAL# (B)(4), IMPLANTED: 2009-(B)(6), PRODUCT TYPE EXTENSION. PRODUCT ID 37085-60, SERIAL# (B)(4), IMPLANTED: 2009-(B)(6), PRODUCT TYPE EXTENSION. PRODUCT ID 3387S-40, LOT# V298925, IMPLANTED: 2009-(B)(6), PRODUCT TYPE LEAD. PRODUCT ID 3387S-40, LOT# V298925, IMPLANTED: 2009-(B)(6), PRODUCT TYPE LEAD. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE LOW OUT OF RANGE IMPEDANCE READINGS. THEY WERE AT THE POST-OPERATION AFTER A PRIMARY CELL DEVICE WAS REPLACED AND THEY HAD NOTED 9/11 PAIR HAD IMPEDANCE OF 39 OHMS. THE SHORT WAS FIRST NOTICED ON THE DATE OF THIS REPORT. IMPEDANCES WERE C/11-605, C/9-610, C/8-1285, C/10-867, 8/9-1323, 8/10-1683, 8/11-1315, 9/10-785, 9/11-39, 10/11-789. THEY WERE UNSURE OF THERAPY STATUS BUT THE PATIENT HAD BEEN PROGRAMMED TO 9-10 PRIOR TO THE REPLACEMENT ON THE DATE OF THIS REPORT. NO OUTCOME WAS PROVIDED. FURTHER FOLLOW-UP IS BEING CONDUCTED. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE IMPEDANCE ISSUE WAS RESOLVED TO SOME DEGREE. THE PATIENT WAS PROGRAMMED BY THE DOCTOR AND DISCHARGED FROM THE HOSPITAL ON (B)(6) 2014. THE CAUSE OF THE IMPEDANCES WAS UNKNOWN. THE REPORTER HAD NOT HEARD FROM THE PATIENT SINCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
50340 ACTIVA STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR MHY MEDTRONIC PUERTO RICO OPERATIONS CO. 37601

Patients

Seq Age Sex Outcome Treatment
1 00066 YR