FDA Adverse Event Malfunction Summary report: N

TEGO CONNECTOR

MDR report key: 4437857 · Received December 15, 2014

Report

Report Number
2025816-2014-00151
Event Type
Malfunction
Date Received
December 15, 2014
Date of Event
September 27, 2014
Report Date
October 13, 2014
Manufacturer
ICU MEDICAL, INC.
Product Code
FPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: CODE: OTHER - A REVIEW OF THE MFG. LOT BUILD DATABASE FOR THE REPORTED LOT# 2820490 (MFG. DATE 02/2014) SHOWS (B)(4) UNITS WERE MFG., TESTED, INSPECTED AND RELEASED. THERE WERE NO EXCEPTION DOCUMENTS GENERATED DURING THE LOT BUILD. CONCLUSION: THE INVOLVED DEVICES WERE NOT RETURNED FOR ANALYSIS AND CONFIRMATION. THE REPORTED INFORMATION DID NOT IDENTIFY ANY VISUAL DEFECTS, COMPONENT DAMAGES WITH THE TEGO CONNECTORS. THE EXACT CAUSE(S) OF THE REPORTED EVENT ARE UNKNOWN AT THIS TIME.

Description of Event or Problem · 1

INT'L COMPLAINT ((B)(4)) RECEIVED REPORTING AIR LEAKS WITH USE OF DIALYSIS SET-UP CONSISTING OF PALINDROME CATHETER; BLOODLINE BAXTER/GAMBRO/HOSPAL ARTIS REF. (B)(4) TEGO CONNECTORS. THE INITIAL INFORMATION (AS TRANSLATED) REPORTS "PRESENCE OF AIR INSIDE THE ARTERIAL BRANCH OF THE PALINDROME CATHETER, WITHOUT CRACK. THE TEGO CAPS WERE SUSPECTED.". THE TEGO CONNECTORS WERE REMOVED AND DISCARDED. THERE WERE NO REPORTED ADVERSE PT CONSEQUENCES. THE TEGO MANUFACTURER HAS REQUESTED ADDITIONAL EVENT INFORMATION AND REQUEST FOR SAME LOT SAMPLES. AS OF THE DATE OF THIS REPORT THERE HAS BEEN NO RESPONSE. INVESTIGATION: A REVIEW OF THE MFG. LOT BUILD DATABASE FOR THE REPORTED LOT# 2820490 (MFG DATE 02/2014) SHOWS (B)(4) UNITS WERE MFG., TESTED, INSPECTED AND RELEASED. THERE WERE NO EXCEPTION DOCUMENTS GENERATED DURING THE LOT BUILD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
821807 TEGO CONNECTOR TEGO CONNECTOR FPA ICU MEDICAL, INC. D1000 2820490

Patients

Seq Age Sex Outcome Treatment
1 NI PALINDROME DIALYSIS TUNNELED CATHETER| BAXTER/GAMBO BLOOD LINES