FDA Adverse Event Malfunction Summary report: N

ENDURO MENISCAL COMPONENT F2 10MM

MDR report key: 4436278 · Received January 14, 2015

Report

Report Number
3005673311-2015-00005
Event Type
Malfunction
Date Received
January 14, 2015
Report Date
January 14, 2015
Manufacturer
AESCULAP AG AND CO. KG
Product Code
KRO
PMA / PMN Number
K101815
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PMA/510 (K) #: K101815/K120955. US REPORTING AGENT NOTIFIED ON (B)(4) 2015. MFG SITE EVAL: ENDURO COMPONENT WAS FOUND TO BE FRACTURED BELOW THE CONE. THE FRAGMENT WITH THE CONE IS FIXED FIRMLY IN THE FEMUR COMPONENT AND THEREFORE, THE FEMUR COMPONENT HAS TO BE REMOVED. THE MFG DOCUMENTS OF ALL INVOLVED COMPONENTS WERE CHECKED AND ALL WERE FOUND TO BE FREE OF ANY DEVIATIONS. THE FRACTURE SURFACE EXHIBITS A FATIGUE FRACTURE WHICH INITIATED THE POSTERIOR (POSTERO-LATERAL) SIDE AND PROTRUDED ANTERIORLY (ANTERO-MEDIAL). FROM THE MARKINGS ON THE FEMUR COMPONENT, IT CAN BE SEEN THAT THE PT WAS EXERTING A HIGH MOMENT TOWARD HYPEREXTENSION. THIS MOST LIKELY CAUSED THE CRACK INITIATION ON THE POSTERO-LATERAL EDGE OF THE ROTATION AXIS.

Description of Event or Problem · 1

COUNTRY OF COMPLAINT: (B)(6). THE PT HAS A POLYNEUROPATHY. AFTER THE SURGERY, THE PT WAS APPARENTLY ONLY MOBILE WITH WALKING AIDS. ON (B)(6) 2014, THE PT WENT TO DR WITH ACUTE INSTABILITY. HE NOTED METALLIC NOISES. A REVISION WAS CARRIED-OUT AND THE AXIS WAS FOUND TO BE BROKEN. INVOLVED COMPONENTS: NB018K LOT # 51800517/ ENDURO FEMORAL COMPONENT CEMENTED F2R. NR410K LOT # 51774154/ FEMUR EXTENS. STEM 5 DEGREE D20X117MM CEM.LESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
34307 ENDURO MENISCAL COMPONENT F2 10MM KNEE PROTHESIS KRO AESCULAP AG AND CO. KG NR880M 51791495

Patients

Seq Age Sex Outcome Treatment
1 Other