ENDURO MENISCAL COMPONENT F2 10MM
Report
- Report Number
- 3005673311-2015-00005
- Event Type
- Malfunction
- Date Received
- January 14, 2015
- Report Date
- January 14, 2015
- Manufacturer
- AESCULAP AG AND CO. KG
- Product Code
- KRO
- PMA / PMN Number
- K101815
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
PMA/510 (K) #: K101815/K120955. US REPORTING AGENT NOTIFIED ON (B)(4) 2015. MFG SITE EVAL: ENDURO COMPONENT WAS FOUND TO BE FRACTURED BELOW THE CONE. THE FRAGMENT WITH THE CONE IS FIXED FIRMLY IN THE FEMUR COMPONENT AND THEREFORE, THE FEMUR COMPONENT HAS TO BE REMOVED. THE MFG DOCUMENTS OF ALL INVOLVED COMPONENTS WERE CHECKED AND ALL WERE FOUND TO BE FREE OF ANY DEVIATIONS. THE FRACTURE SURFACE EXHIBITS A FATIGUE FRACTURE WHICH INITIATED THE POSTERIOR (POSTERO-LATERAL) SIDE AND PROTRUDED ANTERIORLY (ANTERO-MEDIAL). FROM THE MARKINGS ON THE FEMUR COMPONENT, IT CAN BE SEEN THAT THE PT WAS EXERTING A HIGH MOMENT TOWARD HYPEREXTENSION. THIS MOST LIKELY CAUSED THE CRACK INITIATION ON THE POSTERO-LATERAL EDGE OF THE ROTATION AXIS.
COUNTRY OF COMPLAINT: (B)(6). THE PT HAS A POLYNEUROPATHY. AFTER THE SURGERY, THE PT WAS APPARENTLY ONLY MOBILE WITH WALKING AIDS. ON (B)(6) 2014, THE PT WENT TO DR WITH ACUTE INSTABILITY. HE NOTED METALLIC NOISES. A REVISION WAS CARRIED-OUT AND THE AXIS WAS FOUND TO BE BROKEN. INVOLVED COMPONENTS: NB018K LOT # 51800517/ ENDURO FEMORAL COMPONENT CEMENTED F2R. NR410K LOT # 51774154/ FEMUR EXTENS. STEM 5 DEGREE D20X117MM CEM.LESS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 34307 | ENDURO MENISCAL COMPONENT F2 10MM | KNEE PROTHESIS | KRO | AESCULAP AG AND CO. KG | NR880M | 51791495 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |