FDA Adverse Event Injury Summary report: N

TVC INSIGHT CATHETER

MDR report key: 4431960 · Received January 16, 2015

Report

Report Number
3004722468-2014-00003
Event Type
Injury
Date Received
January 16, 2015
Date of Event
December 16, 2014
Report Date
January 15, 2015
Manufacturer
INFRAREDX INC.
Product Code
OGZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CATHETER WAS RETURNED AND WAS INSPECTED BOTH VISUALLY AND FUNCTIONALLY. THERE WERE NO VISUAL EVIDENCE OF FAILURE OR MIS-USE. THE CATHETER HAD NO VISUAL SIGNS OF KINKING OR BREAKAGE. CATHETER WAS SHOWN TO BE FULLY FUNCTIONALLY, TACKING WELL THROUGH A SIMULATED ARCH AND PRODUCING BOTH AN IVUS IMAGE AS WELL AS NIR DATA. NO PROBLEM FOUND.

Description of Event or Problem · 1

PERCUTANEOUS CORONARY INTERVENTION: SUCCESSFUL DEPLOYMENT OF 3.5 X 18MM RESOLUTE DES TO MID-LCX, 75% TO 0% WITH EXCELLENT ANGIOGRAPHIC RESULT AND TIMI 3 FLOW. DISSECTION WAS NOTED PROXIMAL TO THE MID-STENT, SUCCESSFULLY COVERED WITH STENT IN PROXIMAL LCX. SUCCESSFUL DEPLOYMENT OF 3.5 X 18MM RESOLUTE DES TO PROXIMAL LCX, 75% TO 0% WITH EXCELLENT ANGIOGRAPHIC RESULTS AND TIMI 3 FLOW. PATIENT UNDERWENT A COMPLEX INTERVENTION AND HAD CHEST PAIN DURING THE PROCEDURE. NO COMPLICATIONS WERE NOTED DURING THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
40955 TVC INSIGHT CATHETER NIRS IVUS CORONARY IMAGING CATHETER OGZ INFRAREDX INC. TVC-C195-22 D01234

Patients

Seq Age Sex Outcome Treatment
1 58 YR Life Threatening| R