FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 4429573 · Received January 17, 2015

Report

Report Number
2939301-2015-01715
Event Type
Malfunction
Date Received
January 17, 2015
Report Date
January 5, 2015
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING THAT THE SUBJECT METER (ONETOUCH VERIO IQ) READ INACCURATELY HIGH COMPARED TO ANOTHER DEVICE (ONETOUCH ULTRA2). THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF ¿189, 153 AND 169 MG/DL¿ WITH THE SUBJECT METER AND ¿143, 102 AND 125 MG/DL¿ ON ANOTHER DEVICE, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR ACCURACY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
42564 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL

Patients

Seq Age Sex Outcome Treatment
1 65 YR