FDA Adverse Event Other Summary report: N

AIR SHIELD

MDR report key: 442868 · Received February 5, 2003

Report

Report Number
442868
Event Type
Other
Date Received
February 5, 2003
Date of Event
April 1, 2002
Report Date
December 16, 2002
Manufacturer
HILL ROM/AIR SHIELD
Product Code
FMT
Product Problem
Yes
Report Source
User Facility report
Reporter Location
TN, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

TEMPERATURE CONTROL FAILED ON THE SOLID-STATE RELAY HEATING ELEMENTS. THE DEVICE WOULD OVERHEAT, OR NOT HEAT, OR HEAT CONSTANTLY WITH NO ALARMS IN ANY OF THESE SITUATIONS. THE DEVICE SHOULD TURN ON OR OFF DEPENDING ON THE PT'S TEMPERATURE DETECTED BY A PROBE. IF THE SENSOR DIDN'T DETECT CERTAIN TEMPERATURES, THE PT WOULD POTENTIALLY BE COMPROMISED BY DECREASED OR INCREASED TEMPERATURES. THE FACILITY WAS TOLD BY THE MFR THAT THE SERVICE MANUAL INSTRUCTED THEM TO REPLACE THE SOLID-STATE RELAY ON A ROUTINE BASIS BUT THIS REQUIREMENT WAS NOT FOUND BY THE FACILITY IN THE BOOKLET PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AIR SHIELD OPEN BED INFANT WARMER FMT HILL ROM/AIR SHIELD RW82-1 *

Patients

Seq Age Sex Outcome Treatment
1 * Other