- Manufacturer
- HILL ROM/AIR SHIELD
- Product Code
- FMT
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- TN, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
TEMPERATURE CONTROL FAILED ON THE SOLID-STATE RELAY HEATING ELEMENTS. THE DEVICE WOULD OVERHEAT, OR NOT HEAT, OR HEAT CONSTANTLY WITH NO ALARMS IN ANY OF THESE SITUATIONS. THE DEVICE SHOULD TURN ON OR OFF DEPENDING ON THE PT'S TEMPERATURE DETECTED BY A PROBE. IF THE SENSOR DIDN'T DETECT CERTAIN TEMPERATURES, THE PT WOULD POTENTIALLY BE COMPROMISED BY DECREASED OR INCREASED TEMPERATURES. THE FACILITY WAS TOLD BY THE MFR THAT THE SERVICE MANUAL INSTRUCTED THEM TO REPLACE THE SOLID-STATE RELAY ON A ROUTINE BASIS BUT THIS REQUIREMENT WAS NOT FOUND BY THE FACILITY IN THE BOOKLET PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AIR SHIELD | OPEN BED INFANT WARMER | FMT | HILL ROM/AIR SHIELD | RW82-1 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |