FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4419678 · Received January 14, 2015

Report

Report Number
3004209178-2015-84409
Event Type
Malfunction
Date Received
January 14, 2015
Date of Event
December 23, 2014
Report Date
December 23, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FINDINGS: PUMP RECEIVED WITH SEVERELY CRACKED AND BLEEDING LCD GLASS. UNABLE TO PERFORM IDLE CURRENT, RUN CURRENT, SELF TEST, OFF NO POWER, A21 ERROR, DISPLACEMENT, BASIC OCCLUSION, OCCLUSION, PRIME/ A33 AND NO DELIVERY TEST DUE TO LCD DAMAGED. ALSO RECEIVED WITH SEVERELY SCRATCHED DISPLAY WINDOW, CASE DAMAGED, BROKEN BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Description of Event or Problem · 1

CUSTOMER CALLED TO REPORT DAMAGE TO THE INSULIN PUMP. IT WAS REPORTED THAT THE CUSTOMER WAS IN A MOTORCYCLE ACCIDENT AND LANDED ON THE INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE READING WAS 125 MG/DL. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
32041 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 29 YR