FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 4418634 · Received January 14, 2015

Report

Report Number
2531779-2015-01279
Event Type
Malfunction
Date Received
January 14, 2015
Report Date
December 29, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 03/17/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 02/25/2015 WITH THE FOLLOWING FINDINGS: ¿PUMP NOT PRIMED¿ WARNINGS WITH ZERO FORCE AND ¿NO CARTRIDGE DETECTED¿ WARNINGS OBSERVED IN THE BLACK BOX. DURING THE REWIND STEP THE PUMP GAVE ¿076-0001¿ ALARM. THE PUMP WAS OPENED; INTERNAL MOTOR FAILURE FOUND. NO DAMAGE FOUND TO FORCE SENSOR CIRCUIT. BATTERY CAP UNAVAILABLE FOR TESTING A TEST CAP WAS USED FOR TESTING PURPOSES.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
33195 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1