ILAB ULTRASOUND IMAGING SYSTEM
Report
- Report Number
- 2134265-2014-08269
- Event Type
- Malfunction
- Date Received
- January 13, 2015
- Date of Event
- December 15, 2014
- Report Date
- December 15, 2014
- Manufacturer
- BOSTON SCIENTIFIC - FREMONT (CE)
- Product Code
- OBJ
- PMA / PMN Number
- K051679
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4).
DEVICE EVALUATED BY MANUFACTURER: THE COMPLAINT DEVICE WAS RETURNED FOR ANALYSIS. THE PRODUCT WAS RECEIVED IN GOOD CONDITION. THE UNIT MEETS SPECIFICATIONS FOR FUNCTIONAL TEST. IN ADDITION, THE UNIT SUCCESSFULLY UNDERWENT A CONTINUOUS BURN-IN OVERNIGHT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE ROOT CAUSE IS NOT CONFIRMED AS THERE WAS NO EVIDENCE OF THE ALLEGED ISSUE OR ANY ANOMALIES WHICH COULD HAVE CONTRIBUTED TO THE REPORTED DIFFICULTY. (B)(4).
SAME CASE AS: 2134265-2014-08449, 2134265-2014-08450. IT WAS REPORTED THAT AUTOMATIC PULLBACK FAILURE OCCURRED. IT WAS NOTED THAT THE MOTOR DRIVE UNIT (MDU) OF AN ILAB ULTRASOUND IMAGING SYSTEM WAS NOT PULLING BACK. THE LCD DISPLAY WAS NORMAL. THE MDU WAS THEN CHANGED WITH A DIFFERENT MDU AND THEN IT WORKED FINE. NO PATIENT INVOLVEMENT WAS REPORTED.
SAME CASE AS: 2134265-2014-08449, 2134265-2014-08450. IT WAS REPORTED THAT AUTOMATIC PULLBACK FAILURE OCCURRED. IT WAS NOTED THAT THE MOTOR DRIVE UNIT (MDU) OF AN ILAB ULTRASOUND IMAGING SYSTEM WAS NOT PULLING BACK. THE LCD DISPLAY WAS NORMAL. THE MDU WAS THEN CHANGED WITH A DIFFERENT MDU AND THEN IT WORKED FINE. NO PATIENT INVOLVEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 28226 | ILAB ULTRASOUND IMAGING SYSTEM | CATHETER, ULTRASOUND, INTRAVASCULAR | OBJ | BOSTON SCIENTIFIC - FREMONT (CE) | H749ILAB120CART0 | 0000004415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |