FDA Adverse Event Other Summary report: N

PALCO LABS, INC. PEDIATRIC ADJUSTABLE SENSOR

MDR report key: 440169 · Received January 27, 2003

Report

Report Number
2937177-2003-00001
Event Type
Other
Date Received
January 27, 2003
Date of Event
November 20, 2002
Report Date
January 9, 2003
Manufacturer
PALCO LABORATORIES, INC.
Product Code
DQA
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

ON 12/10/2002 ONE OF PALCO'S PRODUCT MANAGERS NOTIFIED THE QA MANAGER THAT A PULSE OXIMETER CUSTOMER REPORTED THE PALCO PULSE OXIMETER MODEL 400 USED WITH A PEDIATRIC ADJUSTABLE SENSOR ON A CHILD WAS CAUSING "BLISTER" THAT COULD BE 'BURNS' ON THEIR FINGERS AND TOES (SENSOR SITES). WHEN THE PRODUCT MANAGER RETURNED THE CALL THEY SPOKE TO THE ATTENDING RESPIRATORY THERAPIST WHO REPORTED THAT THE MARKS HAD BEEN APPEARING OVER A PERIOD OF 3 WEEKS DURING WHICH TIME THE AT-HOME PATIENT WAS BEING CONTINUOUSLY MONITORED FOR %SPO2. THE CUSTOMER REPORTED THAT THE CAREGIVER CHANGED THE SENSOR SITE EVERY 5 TO 8 HOURS. THE RESPIRATORY THERAPIST INDICATED THAT THE PATIENT IS ON A NUMBER OF MEDICATIONS AND SUSPECTS THE PATIENT'S SKIN IS VERY SENSITIVE. A SUBSEQUENT DISCUSSION WITH THE CUSTOMER AND PALCO'S QA MANAGER REVEALED THAT THE SKIN IRRITATION APPEARS ON THE TOP AND BOTTOM OF THE SENSOR SITE ONLY WHERE THE LENSES TOUCH THE SKIN. THE CAREGIVER ALTERS THE SENSOR BY CUTTING THE STRAP AND USES A VELCRO OR 'POSEY-WRAP' TO SECURE THE SENSOR AND LEAVES THE SENSOR ON THE SITE FOR UP TO 8 HOURS. PALCO SENSORS HAVE BEEN USED ON THE PATIENT FOR OVER A YEAR WITHOUT PREVIOUS PROBLEMS WITH SKIN IRRITATION. THE IRRITATION WAS NOT CONSIDERED SERIOUS BY THE CUSTOMER ONLY WORTH NOTING AND QUESTIONING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PALCO LABS, INC. PEDIATRIC ADJUSTABLE SENSOR PULSE OXIMETER SENSOR DQA PALCO LABORATORIES, INC. * *

Patients

Seq Age Sex Outcome Treatment
1 12 YR Other PALCO LABS, INC. PULSE OXIMETER MODEL 400.