FDA Adverse Event Summary report: N

COLLIMATOR ACCESSORY LATCH

MDR report key: 43973 · Received October 21, 1996

Report

Report Number
MW1010122
Date Received
October 21, 1996
Manufacturer
MITSUBISHI INTL CORP.
Product Code
IYC
Report Source
Voluntary report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

COLLIMATOR ACCESSORY LATCH FAILED DURING GANTRY MOVEMENT CAUSING THE WEDGE MOUNT TRAY AND BLOCK MOUNT TRAY TO FALL TO THE FLOOR. A PT WAS ON THE TREATMENT TABLE AND THE TRAYS BARELY MISSED THE PT. A SVC BULLETIN WAS ISSUED BY MFR REGARDING A PROBLEM OF THIS NATURE. IT CALLS FOR THE PARTS TO BE INSPECTED AT THE TIME OF A PM AS A MINIMUM. A PM WAS DONE SEVEN DAYS PRIOR TO THIS INCIDENT. THIS BULLETIN WAS ISSUED 12/28/89 AND THERE HAS BEEN NO ATTEMPT THAT RPTR IS AWARE OF TO CORRECT THE DEFECT. (*)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLIMATOR ACCESSORY LATCH COLLIMATOR IYC MITSUBISHI INTL CORP. EXL-8 *

Patients

Seq Age Sex Outcome Treatment
1 *