FDA Adverse Event
Summary report: N
COLLIMATOR ACCESSORY LATCH
MDR report key: 43973
·
Received October 21, 1996
Report
- Report Number
- MW1010122
- Date Received
- October 21, 1996
- Manufacturer
- MITSUBISHI INTL CORP.
- Product Code
- IYC
- Report Source
- Voluntary report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
COLLIMATOR ACCESSORY LATCH FAILED DURING GANTRY MOVEMENT CAUSING THE WEDGE MOUNT TRAY AND BLOCK MOUNT TRAY TO FALL TO THE FLOOR. A PT WAS ON THE TREATMENT TABLE AND THE TRAYS BARELY MISSED THE PT. A SVC BULLETIN WAS ISSUED BY MFR REGARDING A PROBLEM OF THIS NATURE. IT CALLS FOR THE PARTS TO BE INSPECTED AT THE TIME OF A PM AS A MINIMUM. A PM WAS DONE SEVEN DAYS PRIOR TO THIS INCIDENT. THIS BULLETIN WAS ISSUED 12/28/89 AND THERE HAS BEEN NO ATTEMPT THAT RPTR IS AWARE OF TO CORRECT THE DEFECT. (*)
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLIMATOR ACCESSORY LATCH | COLLIMATOR | IYC | MITSUBISHI INTL CORP. | EXL-8 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |