FDA Adverse Event Malfunction Summary report: N

STEALTH STATION SYSTEM

MDR report key: 439603 · Received January 23, 2003

Report

Report Number
1723170-2003-00001
Event Type
Malfunction
Date Received
January 23, 2003
Date of Event
December 1, 2002
Report Date
January 23, 2003
Manufacturer
MEDTRONIC SURGICAL NAVIGATION TECHNOLOGIES
Product Code
HAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE STEALTHSTATION SOFTWARE TARGET CENTRIC GUIDANCE VIEW ASSISTS THE SURGEON IN DIRECTING AN IMAGE GUIDE INSTRUMENT ALONG A SURGICAL PLAN TOWARD A SPECIFIC TARGET. THIS VIEW IS MOST COMMONLY USED FOR IMAGE GUIDED NEEDLE BIOSY PROCEDURES. THE TARGET CENTRIC GUIDANCE VIEW MONITORS THE POSITION OF THE INSTRUMENT'S TIP ALONG THE SURGICAL PLAN AS THE INSTRUMENT IS ADVANCED. THE POSITION ALONG THE PLAN IS REPRESENTED BY A BLUE ARROW THAT MOVES DOWN THE RIGHT EDGE OF THE GUIDANCE VIEW AS THE INSTRUMENT IS ADVANCED. WHEN THE BLUE ARROW REACHES THE BORDER OF THE RED AREA, THE TOP OF THE INSTRUMENT HAS REACHED ITS PREDEFINED TARGET. HOWEVER, IN THE STEALTHSTATION SOFTWARE TARGET CENTRIC GUIDANCE VIEW VERSIONS 4.1.2 AND 4.1.3 THE INDICATOR ARROW REVERSES DIRECTION, INDICATING WITHDRAWAL AS THE INSTRUMENT ADVANCES. THE DISTANCE THE INDICATOR ARROW WITHDRAWS IS THE EQUIVALENT TO THE DISTANCE BY WHICH THE TARGET HAS BEEN OVERSHOT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STEALTH STATION SYSTEM STEREOTACTIC DEVICE HAW MEDTRONIC SURGICAL NAVIGATION TECHNOLOGIES 960-095 NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other