ACTIVA
Report
- Report Number
- 3004209178-2015-00259
- Event Type
- Malfunction
- Date Received
- January 7, 2015
- Report Date
- December 15, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- MHY
- PMA / PMN Number
- P960009
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4).
IT WAS REPORTED THE PATIENT EXPERIENCED A ¿BURNING SENSATION AT HER IMPLANTABLE NEUROSTIMULATOR (INS) SITE WHEN THE VOLTAGE WAS INCREASED TO 3 VOLTS OR ABOVE.¿ THE PATIENT¿S VOLTAGE WAS SET TO BELOW 3 VOLTS AT THE TIME OF REPORT ¿IN ORDER TO RELIEF SUCH SENSATION.¿ THE PATIENT WAS REPORTEDLY ¿RECEIVING EFFECTIVE THERAPY;¿ HOWEVER, SHE ¿COULDN¿T OPTIMIZE THE THERAPEUTIC EFFECT BY INCREASING THE VOLTAGE BECAUSE OF THE BURNING SENSATION.¿ IT WAS UNKNOWN WHETHER THE ISSUE WAS CAUSED BY THE INS AS OF FOUR DAYS AFTER INITIAL REPORT. A SUPPLEMENTAL REPORT WILL BE FILED IF ADDITIONAL INFORMATION IS RECEIVED.
ADDITIONAL INFORMATION REPORTED ¿THE SENSATION AS ONLY PRESENT WHEN THE STIMULATOR WAS ON¿ AND THAT THE BURNING SENSATION ¿DID NOT APPEAR ALL THE DAY.¿ WHEN THE PATIENT¿S STIMULATION VOLTAGE WAS INCREASED ABOVE THREE VOLTS, IT NOTABLY ¿TOOK TIME FOR THE PATIENT TO HAVE THE FEELING.¿ IMPEDANCE TESTING WAS PERFORMED AND NOTED ¿IMPEDANCES WERE SLIGHTLY OUT OF RANGE¿ AND THAT THERE WAS ¿AN OPEN CIRCUIT.¿ THE MONOPOLAR IMPEDANCE VALUES FOR ELECTRODES 8, 9, AND 10 WERE FOUND TO BE 2532, 2180, AND 2303 OHMS RESPECTIVELY, WHILE THE BIPOLAR IMPEDANCE VALUES FOR ELECTRODE PAIRS 8-10, 8-11, AND 9-11 WERE FOUND TO BE 4050, 4161, AND 4018 OHMS RESPECTIVELY. THE PATIENT HAD NOT REPORTED ANY FALLS OR TRAUMATIC EVENTS AS OF THE TIME OF FOLLOW-UP.
ADDITIONAL INFORMATION RECEIVED REPORTED THAT IMPEDANCE WAS SLIGHTLY OUT OF RANGE ON THE RIGHT LEAD AND THERE WAS AN OPEN CIRCUIT. THE IMPEDANCES WERE GREATER THAN 2,000 OHMS FOR MONOPOLAR AND GREATER THAN 4 ,000 OHMS FOR BIPOLAR WHEN MEASURED AT 1.5V. THE FOLLOWING OUT OF RANGE IMPEDANCES WERE MEASURED: C-8 = 2532 OHMS, C-9 = 2180 OHMS, C-10 = 2303 OHMS, 8-10 = 4050 OHMS, 8-11 = 4161 OHMS, AND 9-11 = 4018 OHMS. OTHER ELECTRODE PAIRS WERE NOT ABLE TO BE TESTED DUE TO SIDE EFFECTS. THE PATIENT COULD NOT WITHSTAND THE SPASM GENERATED BY OTHER CONTACTS. X-RAYS WERE DONE AND THERE WAS NO APPARENT DAMAGE OR BREAKAGE IN THE LEAD OR EXTENSIONS. AT THE TIME OF THIS REPORT, THE PATIENT WAS USING BIPOLAR SETTINGS INSTEAD OF UNIPOLAR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 13914 | ACTIVA | STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR | MHY | MEDTRONIC PUERTO RICO OPERATIONS CO. | 37601 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |