FDA Adverse Event Other Summary report: N

BERCHTOLD TELETOM

MDR report key: 439248 · Received January 17, 2003

Report

Report Number
1220685-2003-00001
Event Type
Other
Date Received
January 17, 2003
Date of Event
December 19, 2002
Report Date
January 13, 2003
Manufacturer
BERCHTOLD CORP.
Product Code
FQO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

DURING A SURGICAL PROCEDURE, AN O.R. NURSE WENT TO RE-POSITION A TELETOM (BOOM), WHICH WAS FIVE FEET AWAY FROM THE OPERATING TABLE. WHEN THE BOOM ROTATED TO AN INTERNAL STOP, THE NURSE ATTEMPTED TO FORCEFULLY OVERCOME THE STOP BY EXERTING ADDITIONAL PRESSURE. THE COMBINATION OF LATERAL AND VERTICAL FORCES LIFTED THE EQUIPMENT COLUMNS & SHELVES OFF THE BOOM, CAUSING THEM TO FALL APPROXIMATELY 2 FEET TO THE FLOOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BERCHTOLD TELETOM GAS & EQUIPMENT COLUMN FQO BERCHTOLD CORP. 700 NA

Patients

Seq Age Sex Outcome Treatment
1 *