G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-00049
- Event Type
- Malfunction
- Date Received
- January 6, 2015
- Date of Event
- December 9, 2014
- Report Date
- December 9, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4). SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT RECEIVER DEVICE WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER DEVICE(9438-01/64EKE), USED WITH THE COMPLAINT RECEIVER DEVICE, WAS RETURNED ON 02/09/2015. THE RETURNED COMPLAINT TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE TEST FAILED, CONFIRMING THE REPORTED EVENT OF A PERMANENT OUT OF RANGE SIGNAL. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE TRANSMITTER. A CAPA HAS BEEN INITIATED FOR THIS ISSUE.
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO REPORT A PERMANENT OUT OF RANGE SIGNAL ON (B)(6) 2014. PATIENT WAS ADVISED TO RESET THE DEVICE AND THE RESET WAS UNSUCCESSFUL FOR THE REPORTED ISSUE. AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 7805 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT20649-BLU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |