FDA Adverse Event Malfunction Summary report: N

AIRSHIELDS

MDR report key: 438707 · Received January 20, 2003

Report

Report Number
MW1027320
Event Type
Malfunction
Date Received
January 20, 2003
Date of Event
January 19, 2003
Report Date
January 20, 2003
Manufacturer
HILL-ROM AIR SHIELDS INC
Product Code
FMZ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
KY, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

A PT FELL OUT OF THE ACCESS DOOR OF AN INCUBATOR. BIO-MED EVALUATED THE DEVICE AND FOUND THE FOLLOWING: THREE OF THE LATCHES WERE FOUND TO REQUIRE THE OPERATOR TO PUSH ON THE DOOR TO FULLY ENGAGE THE LATCH, A SIMPLE FLIP OF THE DOOR WOULD ONLY HALFWAY ENGAGE THE DOOR LATCH. THESE WERE REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AIRSHIELDS INFANT INCUBATOR FMZ HILL-ROM AIR SHIELDS INC C100/200-2 *

Patients

Seq Age Sex Outcome Treatment
1 *