STOCKERT EP-SHUTTLE RADIO FREQUENCY GENERATOR
Report
- Report Number
- 9612355-2015-00001
- Event Type
- Injury
- Date Received
- January 6, 2015
- Date of Event
- September 26, 2014
- Report Date
- December 23, 2014
- Manufacturer
- STOCKERT GMBH
- Product Code
- LPB
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THE ¿SUSPECTED MEDICAL DEVICE¿ REPORTED IN SECTION D OF THIS REPORT IS NOT MARKETED IN USA OR APPROVED BY THE FDA. HOWEVER, IT IS BEING REPORTED AS BIOSENSE WEBSTER CONSIDERS THIS AS A SIMILAR DEVICE TO STOCKERT 70 RF GENERATOR APPROVED UNDER PMA # P990071. THE HARDWARE INVESTIGATION HAS BEGUN BUT IT HAS NOT BEEN COMPLETED AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED. (B)(4).
(B)(4). IT WAS REPORTED THAT A FISTULA APPEARED ABOUT 4 WEEKS AFTER THE PROCEDURE. THE DEVICE WAS EVALUATED AND NO ERROR FOUND. DEVICE IS WITHIN SPECIFICATION. THE DEVICE WAS SUBJECTED TO THE PREVENTIVE MAINTENANCE, SAFETY AND FUNCTIONAL TESTING AND ALL TESTS PASSED. NO MALFUNCTION FOUND ON DEVICE. THE DEVICE HISTORY RECORD WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES.
IT WAS REPORTED THAT A PATIENT, UNDERWENT AN ATRIAL FIBRILLATION PROCEDURE WITH A STOCKERT EP-SHUTTLE RADIO FREQUENCY GENERATOR AND 4 WEEKS AFTER THE PROCEDURE SUFFERED AN ESOPHAGEAL FISTULA WHICH REQUIRED OPERATION. THE PATIENT¿S MEDICAL HISTORY IS UNKNOWN. NO FURTHER INFORMATION IS KNOWN REGARDING THE PATIENT STATUS. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN CLARIFICATION TO THIS COMPLAINT. HOWEVER, NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 6414 | STOCKERT EP-SHUTTLE RADIO FREQUENCY GENERATOR | SIMILAR DEVICE S7001, PMA # P990071 | LPB | STOCKERT GMBH | 39D-76X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Life Threatening| R |