FDA Adverse Event
Injury
Summary report: N
COTTON TIP WOOD APPLICATOR
MDR report key: 4381399
·
Received December 30, 2014
Report
- Report Number
- 1417592-2014-00127
- Event Type
- Injury
- Date Received
- December 30, 2014
- Date of Event
- December 3, 2014
- Report Date
- December 26, 2014
- Manufacturer
- MEDLINE INDUSTRIES, INC.
- Product Code
- KXG
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE APPLICATOR BROKE IN HALF WHEN INSERTED INTO A SACRAL ABSCESS TO MEASURE THE UNDERMINING OF THE WOUND. THE WOUND HAD THICK, PURULENT DRAINAGE. THE RETAINED HALF OF THE APPLICATOR WAS RETRIEVED IN A PHYSICIAN OFFICE WITHOUT FURTHER INCIDENT. NO SAMPLE WAS RETURNED FOR EVALUATION AND A ROOT CAUSE HAD NOT BEEN DETERMINED. A CAUTION STATEMENT ON THE LABELING INDICATES THAT THIS DEVICE SHOULD NOT BE USED IN PROCEDURES WHICH MAY REQUIRE EXCESSIVE PRESSURE. WE CANNOT RULE OUT THE POSSIBILITY THAT PRESSURE EXERTED ON THE STICK COULD HAVE BEEN A CAUSE OR CONTRIBUTING FACTOR TO THE REPORTED INCIDENT.
Description of Event or Problem · 1
THE WOODEN STICK BROKE AFTER BEING INSERTED INTO A SACRAL WOUND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 859594 | COTTON TIP WOOD APPLICATOR | KXG | MEDLINE INDUSTRIES, INC. | 507239725 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Required Intervention |