SIMPLEXA FLU A/B AND RSV DIRECT ASSAY
Report
- Report Number
- 2023365-2014-00029
- Event Type
- Malfunction
- Date Received
- December 29, 2014
- Date of Event
- January 30, 2014
- Report Date
- January 30, 2014
- Manufacturer
- FOCUS DIAGNOSTICS, INC.
- Product Code
- OCC
- PMA / PMN Number
- K120413
- Removal / Correction Number
- Z-1144-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION OF THE SIMPLEXA FLU A/B RSV DIRECT KIT (MOL2650) SUGGESTED THAT THE ROOT CAUSE FOR THE OCCURRENCE OF FALSE POSITIVE (AND SUB-AMPLIFICATIONS) WERE DUE TO THE LOT TO LOT VARIABILITY OF THE SCORPION PRIMERS. TO MITIGATE THE OCCURRENCE OF FALSE POSITIVES (AND SUB-AMPLIFICATIONS), (B)(4). THESE CORRECTIVE AND PREVENTATIVE ACTIONS ARE EFFECTIVE IN MITIGATING THE EFFECTS POSED BY THE POTENTIAL LOT TO LOT VARIABILITY OF THE SCORPIONS. (B)(4).
THE CUSTOMER REPORTED THAT THEY ARE SEEING FALSE POSITIVE FLU B RESULTS WITH THEIR SAMPLE TESTING. THERE HAVE BEEN FOUR SAMPLES THAT CAME UP LATE FOR FLU B AND UPON REPEAT TESTING WITH ANOTHER REFERENCE KIT, ALL FOUR SAMPLES WERE NEGATIVE FOR FLU B. THE RESULTS WERE REPORTED TO THE DOCTOR. THIS MDR IS FOR THE SECOND OF THE FOUR SAMPLE RESULTS THAT WERE REPORTED. A VOLUNTARY RECALL HAS BEEN PERFORMED RELATED TO THE REPORTED LOT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 855480 | SIMPLEXA FLU A/B AND RSV DIRECT ASSAY | IN-VITRO-DIAGNOTIC TEST, INFLUENZA | OCC | FOCUS DIAGNOSTICS, INC. | MOL2650 | 24732 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |