FDA Adverse Event Malfunction Summary report: N

SIMPLEXA FLU A/B AND RSV DIRECT ASSAY

MDR report key: 4379514 · Received December 29, 2014

Report

Report Number
2023365-2014-00029
Event Type
Malfunction
Date Received
December 29, 2014
Date of Event
January 30, 2014
Report Date
January 30, 2014
Manufacturer
FOCUS DIAGNOSTICS, INC.
Product Code
OCC
PMA / PMN Number
K120413
Removal / Correction Number
Z-1144-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION OF THE SIMPLEXA FLU A/B RSV DIRECT KIT (MOL2650) SUGGESTED THAT THE ROOT CAUSE FOR THE OCCURRENCE OF FALSE POSITIVE (AND SUB-AMPLIFICATIONS) WERE DUE TO THE LOT TO LOT VARIABILITY OF THE SCORPION PRIMERS. TO MITIGATE THE OCCURRENCE OF FALSE POSITIVES (AND SUB-AMPLIFICATIONS), (B)(4). THESE CORRECTIVE AND PREVENTATIVE ACTIONS ARE EFFECTIVE IN MITIGATING THE EFFECTS POSED BY THE POTENTIAL LOT TO LOT VARIABILITY OF THE SCORPIONS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY ARE SEEING FALSE POSITIVE FLU B RESULTS WITH THEIR SAMPLE TESTING. THERE HAVE BEEN FOUR SAMPLES THAT CAME UP LATE FOR FLU B AND UPON REPEAT TESTING WITH ANOTHER REFERENCE KIT, ALL FOUR SAMPLES WERE NEGATIVE FOR FLU B. THE RESULTS WERE REPORTED TO THE DOCTOR. THIS MDR IS FOR THE SECOND OF THE FOUR SAMPLE RESULTS THAT WERE REPORTED. A VOLUNTARY RECALL HAS BEEN PERFORMED RELATED TO THE REPORTED LOT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
855480 SIMPLEXA FLU A/B AND RSV DIRECT ASSAY IN-VITRO-DIAGNOTIC TEST, INFLUENZA OCC FOCUS DIAGNOSTICS, INC. MOL2650 24732

Patients

Seq Age Sex Outcome Treatment
1 UNK