SONOSURG-G2 SET

MDR report key: 4379500 · Received December 24, 2014

Report

Report Number
8010047-2014-01100
Event Type
Malfunction
Date Received
December 24, 2014
Date of Event
November 27, 2014
Report Date
December 28, 2015
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
LFL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP (OMSC) FOR EVAL. THE EVAL CONFIRMED THAT THE TWO FUSES OF THE SUBJECT DEVICE WERE BLOWN AND FET (FIELD-EFFECT TRANSISTOR) OF AMPLIFIER SUBSTRATE WAS DAMAGED. THE MFG RECORDS FOR THIS SERIAL NUMBER INDICATED NO ABNORMALITIES RELATED TO THE REPORTED PHENOMENON. BASED UPON THE EVAL RESULT, THE TWO FUSES WERE BLOWN AND THE SUBJECT DEVICE SHUT DOWN DUE TO THE DAMAGE OF FET. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.

Additional Manufacturer Narrative · 1

OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) PERFORMED A MDR RETROSPECTIVE REVIEW AND OMSC FOUND THAT THIS SUPPLEMENTAL REPORT IS REQUIRED ON DECEMBER 7, 2015. THIS IS A SUPPLEMENTAL REPORT FOR MFR REPORT #8010047-2014-01100 TO PROVIDE DEVICE EVALUATION RESULTS. OLYMPUS PERFORMED FURTHER INVESTIGATION FOR THE CAUSE OF THE FET (FIELD EFFECT TRANSISTOR) DAMAGE AND THE DAMAGED FET HAD EXCESSIVE CURRENT LEAKAGE. THE AMOUNT OF THE LEAKAGE CURRENT DEPENDS ON THE ELECTRICAL CHARACTERISTIC VARIABILITY OF THE FET. OLYMPUS IMPLEMENTED A COUNTERMEASURE FOR THIS PHENOMENON TO REDUCE THE ELECTRICAL CHARACTERISTIC VARIABILITY OF THE FET.

Description of Event or Problem · 1

DURING PROCEDURE, THE SUBJECT DEVICE SHUT DOWN. UPON THE INVESTIGATION, THE FUSE WAS FOUND TO BE BLOWN. THE INTENDED PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
850014 SONOSURG-G2 SET ULTRASONIC SURGICAL SYSTEM LFL OLYMPUS MEDICAL SYSTEMS CORPORATION SONOSURG-G2 NA

Patients

Seq Age Sex Outcome Treatment
1