SONOSURG-G2 SET
Report
- Report Number
- 8010047-2014-01100
- Event Type
- Malfunction
- Date Received
- December 24, 2014
- Date of Event
- November 27, 2014
- Report Date
- December 28, 2015
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- LFL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP (OMSC) FOR EVAL. THE EVAL CONFIRMED THAT THE TWO FUSES OF THE SUBJECT DEVICE WERE BLOWN AND FET (FIELD-EFFECT TRANSISTOR) OF AMPLIFIER SUBSTRATE WAS DAMAGED. THE MFG RECORDS FOR THIS SERIAL NUMBER INDICATED NO ABNORMALITIES RELATED TO THE REPORTED PHENOMENON. BASED UPON THE EVAL RESULT, THE TWO FUSES WERE BLOWN AND THE SUBJECT DEVICE SHUT DOWN DUE TO THE DAMAGE OF FET. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) PERFORMED A MDR RETROSPECTIVE REVIEW AND OMSC FOUND THAT THIS SUPPLEMENTAL REPORT IS REQUIRED ON DECEMBER 7, 2015. THIS IS A SUPPLEMENTAL REPORT FOR MFR REPORT #8010047-2014-01100 TO PROVIDE DEVICE EVALUATION RESULTS. OLYMPUS PERFORMED FURTHER INVESTIGATION FOR THE CAUSE OF THE FET (FIELD EFFECT TRANSISTOR) DAMAGE AND THE DAMAGED FET HAD EXCESSIVE CURRENT LEAKAGE. THE AMOUNT OF THE LEAKAGE CURRENT DEPENDS ON THE ELECTRICAL CHARACTERISTIC VARIABILITY OF THE FET. OLYMPUS IMPLEMENTED A COUNTERMEASURE FOR THIS PHENOMENON TO REDUCE THE ELECTRICAL CHARACTERISTIC VARIABILITY OF THE FET.
DURING PROCEDURE, THE SUBJECT DEVICE SHUT DOWN. UPON THE INVESTIGATION, THE FUSE WAS FOUND TO BE BLOWN. THE INTENDED PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 850014 | SONOSURG-G2 SET | ULTRASONIC SURGICAL SYSTEM | LFL | OLYMPUS MEDICAL SYSTEMS CORPORATION | SONOSURG-G2 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |