SIMPLEXA FLU A/B AND RSV DIRECT ASSAY
Report
- Report Number
- 2023365-2014-00050
- Event Type
- Malfunction
- Date Received
- December 29, 2014
- Date of Event
- January 30, 2014
- Report Date
- January 30, 2014
- Manufacturer
- FOCUS DIAGNOSTICS, INC.
- Product Code
- OCC
- PMA / PMN Number
- K120413
- Removal / Correction Number
- Z-1958-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CUSTOMER REPORTED THE EVENT AGAINST CATALOG # MOL1451, THE DIRECT AMPLIFICATION DISK, WHICH IS USED WITH MOL2650. THE LOT NUMBER OF THE MOL1451 PRODUCT WAS REPORTED AS LOT# 132222. THERE IS NO EXPIRATION DATE FOR DAD DISKS. THE MFR DATE OF THE DAD DISC WAS ON (B)(6) 2013. INVESTIGATION DETERMINED THAT CARRY-OVER CONTAMINATION FROM SUCCESSIVE RUNS CAN OCCUR IF THE END-USER STORES THE DAD (DIRECT AMPLIFICATION DISC) VERTICALLY (I.E. ON EDGE). STORAGE OF THE DAD IN ANY VERTICAL POSITION AFTER INITIAL USE HAS A POTENTIAL OF ALLOWING FLUID TO TRAVEL BACK THROUGH THE CHANNELS AND CONTAMINATE OTHER UNUSED WEDGES. THIS MAY RESULT IN OBTAINING FALSE POSITIVE RESULTS. THE PRODUCT INSERT WARNING AND PRECAUTIONS SECTION HAS BEEN REVISED TO INCLUDE THE FOLLOWING ITEM: "AFTER EACH USE STORE DAD DISCS FLAT WITH THE NUMBERED FOIL SIDE UP". MFR. REF. NUMBER #:(B)(4).
THE CUSTOMER REPORTED THAT THEY BELIEVE THAT THEY HAVE A FLU A AMPLICON ISSUE, POSSIBLY RELATED TO REUSING THE DAD DISCS. THE CUSTOMER TOOK ENVIRONMENTAL SWABS FROM VARIOUS AREAS AROUND THEIR LABORATORY. THE SWABS FROM ALL THREE OF THEIR CYCLERS CAME UP POSITIVE FOR FLU A. THEY HAVE NOT FOUND ANY OTHER AREAS OF CONTAMINATION IN THEIR LABORATORY. THEY STORE THE DAD DISCS VERTICALLY IN-BETWEEN USES. THE CUSTOMER DID NOT FEEL THAT THIS WAS AFFECTING THE PT RESULTS. THE RESULTS WERE NOT REPORTED TO THE DOCTOR. THE PRODUCT INSERT WAS REVISED TO ADVISE THE END-USER THAT "AFTER EACH USE STORE DAD DISCS FLAT WITH THE NUMBERED FOIL SIDE UP".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 855159 | SIMPLEXA FLU A/B AND RSV DIRECT ASSAY | IN-VITRO-DIAGNOSTICS TEST, INFLUENZA | OCC | FOCUS DIAGNOSTICS, INC. | MOL2650 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |