FDA Adverse Event Malfunction Summary report: N

SIMPLEXA FLU A/B AND RSV DIRECT ASSAY

MDR report key: 4379428 · Received December 29, 2014

Report

Report Number
2023365-2014-00051
Event Type
Malfunction
Date Received
December 29, 2014
Date of Event
November 1, 2013
Report Date
November 2, 2013
Manufacturer
FOCUS DIAGNOSTICS, INC.
Product Code
OCC
PMA / PMN Number
K120413
Removal / Correction Number
Z-1958-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION DETERMINED THAT CARRY-OVER CONTAMINATION FROM SUCCESSIVE RUNS CAN OCCUR IF THE END-USER STORES THE DAD (DIRECT AMPLIFICATION DISC) VERTICALLY (I.E. ON EDGE). STORAGE OF THE DAD IN ANY VERTICAL POSITION AFTER INITIAL USE HAS A POTENTIAL OF ALLOWING FLUID TO TRAVEL BACK THROUGH THE CHANNELS AND CONTAMINATE OTHER UNUSED WEDGES. THIS MAY RESULT IN OBTAINING FALSE POSITIVE RESULTS. THE PRODUCT INSERT WARNING AND PRECAUTIONS SECTION HAS BEEN REVISED TO INCLUDE THE FOLLOWING ITEM: "AFTER EACH USE STORE DAD DISCS FLAT WITH THE NUMBERED FOIL SIDE UP." (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY ARE HAVING TOO MANY INVALID INTERNAL CONTROL VALUES ON THE PATIENT SAMPLES. NO RESULTS WERE REPORTED TO THE DOCTOR AS THEY WERE UNABLE TO OBTAIN READINGS. REVIEW OF THE CUSTOMER'S RUN DATABASE FOUND FOUR INVALID SAMPLES IN ONE RUN. THE SAMPLES WERE VALID UPON REPEAT TESTING USING THE SAME LOT OF REAGENTS. THE PRODUCT INSERT WAS REVISED TO ADVISE THE END-USER THAT "AFTER EACH USE STORE DAD DISCS FLAT WITH THE NUMBERED FOIL SIDE UP."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
854717 SIMPLEXA FLU A/B AND RSV DIRECT ASSAY IN-VITRO-DIAGNOSTIC TEST, INFLUENZA OCC FOCUS DIAGNOSTICS, INC. MOL2650 23784

Patients

Seq Age Sex Outcome Treatment
1 UNK