Manufacturer
BENVENUE MEDICAL, INC.
Product Code
OCJ
PMA / PMN Number
K070293
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE WORKING CANNULA REMNANT WAS RETURNED. INVESTIGATION OF THE RETURNED PRODUCT WAS CONDUCTED AND THE REPORTED EVENT WAS CONFIRMED. BASED ON THE ANALYSIS, THE BREAKAGE WAS INITIATED BY TORSIONAL BUCKLING EVIDENT BY THE "BARBER POLE" SHAPE LEFT ON CANNULA PIECE. THE INVESTIGATION HOWEVER COULD NOT REPLICATE THE REPORTED EVENT IN A TESTING ENVIRONMENT. THE INVESTIGATION CONCLUDED THAT THE CAUSE OF THE REPORTED EVENT WAS DUE TO THE USE OF THE ACCESS TOOLS EXCEEDED THE WORKING CANNULA TORSIONAL STRENGTH. IN THIS CASE THE PATIENT PRESENTED AN UNUSUAL DEMANDING HARD BONE CONDITION, AS STATED BY THE SURGEON. REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THIS DEVICE PASSED ALL INSPECTION REQUIREMENTS. THERE WERE NO NON-CONFORMANCES IDENTIFIED FOR THIS LOT THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THERE IS NO INFORMATION THAT INDICATES EVIDENCE OF PRODUCT DEFICIENCY. THIS REPRESENTS AN ISOLATED OCCURRENCE; THEREFORE NO CORRECTIVE ACTION WILL BE IMPLEMENTED. BENVENUE WILL CONTINUE TO MONITOR THESE TYPES OF INCIDENTS AND TAKE THE NECESSARY STEPS TO CORRECT ANY FUTURE FINDINGS.

Description of Event or Problem · 1

PATIENT HAD THORACIC (T6) VERTEBRAL COMPRESSION FRACTURE, HAS BEEN GETTING CHEMOTHERAPY, RADIATION AND TAKING BISPHOSPHONATES FOR CANCER TREATMENT. AS THE SURGEON ACCESSED THE T6 PEDICLE WITH THE WORKING CANNULA, IT WAS APPARENT THAT THE PATIENT'S BONE WAS EXTREMELY HARD. THE SURGEON APPLIED QUITE A BIT OF FORCE TO THE CANNULA WHILE MANIPULATING THE CANNULA TO CHANGE THE DIRECTION. AS A RESULT OF THE EXCESSIVE FORCE, MANIPULATION AND TORQUE, THE WORKING CANNULA WAS BENT AND EVENTUALLY BROKE DURING THE MANEUVER WITH 1-2 CM PROTRUDING POSTERIORLY FROM THE PEDICLE BUT STILL SUBCUTANEOUS. THE SURGEON WAS STILL ABLE TO INJECT THE BONE CEMENT SUCCESSFULLY USING THE BLAZER-C DEVICE COMPLETING THE VERTEBRAL AUGMENTATION PROCEDURE. THE WORKING CANNULA WAS CUT AT THE POSTERIOR BORDER OF THE PEDICLE AFTER FURTHER TRIES TO REMOVE IT, AND A PARTIAL PIECE OF THE CANNULA REMAINS SECURELY IN THE PEDICLE. THE SURGEON DID NOT ALLEGE A DEFICIENCY IN THE PERFORMANCE OF THE DEVICE AND INDICATES NO FURTHER PATIENT EFFECT. IN THE SURGEON'S OPINION, THE BLAZER-C PROCEDURE IS A CLEAR CLINICAL BENEFIT TO THE PATIENT. THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
835259 BLAZER-C VERTEBRAL AUGMENTATION SYSTEM OCJ BENVENUE MEDICAL, INC. 14008001

Patients

Seq Age Sex Outcome Treatment
1 Other