FDA Adverse Event Malfunction Summary report: N

COMPLETE DOUBLE MOIST

MDR report key: 4375914 · Received April 20, 2012

Report

Report Number
3004178847-2012-00013
Event Type
Malfunction
Date Received
April 20, 2012
Date of Event
April 20, 2012
Report Date
April 20, 2012
Manufacturer
ABBOTT MEDICAL OPTICS INC.
Product Code
LPN
PMA / PMN Number
K030092
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE MANUFACTURING RECORDS FOR THE REPORTED LAT WAS PERFORMED; NO DEVIATIONS WERE NOTED AND PRODUCT MET ALL SPECIFICATIONS. MICROBIOLOGY EVALUATION OF THE RETAINED UNIT FROM THE REPORTED LOT SHOWED THE SOLUTION TO BE STERILE. PHYSICO-CHEMICAL ANALYSIS RESULTS OF THE RETURNED UNIT ARE CORRECT AND ALL PARAMETERS ARE WITHIN ESTABLISHED ACCEPTANCE CRITERIA. VISUAL INSPECTION PERFORMED ON THE RETURNED SAMPLE. IT MET SPECIFICATION FOR COMPONENTS INTEGRITY AND PRODUCT LEAKAGE. A YELLOW SUBSTANCE WAS NOTED ON THE THREADS OF ONE BOTTLE WHICH WAS NOT WITHIN SPECIFICATION. THE RESIDUE WAS ANALYZED USED SPECTROSCOPY. THE RESIDUE WAS FOUND TO BE DRIED MULTIPURPOSE SOLUTION WITH THE ADDITIONAL PRESENCE OF BIOLOGICAL MATERIAL (PROTEIN). THE COMPLAINT UNIT WAS RETURNED BACK ALREADY OPENED AND MICROBIOLOGY RESULTS SHOW THAT THE SAMPLE WAS NOT STERILE. THE SECOND BOTTLE PACKAGED WITH THE RETURNED UNIT MET MICROBIOLOGICAL, CHEMISTRY AND VISUAL SPECIFICATIONS. ALL PERTINENT INFORMATION AVAILABLE TO AMO HAS BEEN SUBMITTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

OUR OFFICE IN (B)(4) RECEIVED A REPORT FROM A DISTRIBUTOR THAT A CONSUMER FOUND DIRT AND A SMALL INSECT INSIDE OF A BOTTLE CAP WHEN SHE OPENED THE BOTTLE OF COMPLETE DOUBLE MOIST. THE REPORTER INDICATED THE SAFETY SEAL WAS IN PLACE WHEN THE BOTTLE WAS INITIALLY OPENED. THE BOTTLE WAS PART OF A TWO PACK. THE CONSUMER DID NOT USE THE PRODUCT; NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMPLETE DOUBLE MOIST LPN ABBOTT MEDICAL OPTICS INC. ZK00287

Patients

Seq Age Sex Outcome Treatment
1