FDA Adverse Event Malfunction Summary report: N

CUSTOM DISTAL FEMUR IMPLANT

MDR report key: 4373070 · Received December 24, 2014

Report

Report Number
3004105610-2014-00173
Event Type
Malfunction
Date Received
December 24, 2014
Date of Event
May 15, 2014
Report Date
April 30, 2014
Manufacturer
STANMORE IMPLANTS WORLDWIDE LTD
Product Code
KRO
PMA / PMN Number
K121029
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE MANUFACTURING HISTORY CONFIRMS NON NON-CONFORMANCES WERE IDENTIFIED. THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. BASED ON THE LENGTH OF TIME THE DEVICE HAS BEEN IMPLANTED IT IS LIKELY THAT THE REBUSHING PROCEDURE WAS REQUIRED DUE TO NORMAL WEAR AND TEAR OF THE DEVICE. REQUESTS ARE BEING MADE FOR ADDITIONAL INFORMATION AND A SUPPLEMENTAL REPORT WILL BE PROVIDED IF THIS INFORMATION IS RECEIVED. PLEASE NOTE THAT THIS CUSTOM IMPLANT IS SIMILAR TO THE MET SMILES TOTAL KNEE REPLACEMENT IMPLANT (K120992).

Additional Manufacturer Narrative · 1

A REBUSHING PROCEDURE WAS SUCCESSFULLY PERFORMED WITH NO REPORTED COMPLICATIONS. REBUSHING IS AN ANTICIPATED EVENT DUE TO THE NATURE AND DESIGN OF THE IMPLANT. IT CONSISTS OF THE REPLACEMENT OF POLY COMPONENTS WHICH ARE SUSCEPTIBLE TO WEAR SUCH AS BEARINGS AND BUSHINGS, IN ADDITION TO THE AXLE AND RELATED RETAINING MECHANISMS. THIS COMPLAINT IS BEING CLOSED AND IS BEING TRACKED AND TRENDED.

Description of Event or Problem · 1

IT WAS REPORTED BY THE SURGEON THAT THE PATIENT UNDERWENT A DISTAL FEMUR REPLACEMENT PROCEDURE ON (B)(6) 1999 AND SUBSEQUENTLY UNDERWENT A REBUSHING PROCEDURE ON (B)(6), 2014. REF: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY THE SURGEON THAT THE PATIENT UNDERWENT A DISTAL FEMUR REPLACEMENT PROCEDURE ON (B)(6) 1999 AND SUBSEQUENTLY UNDERWENT A REBUSHING PROCEDURE ON (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
851792 CUSTOM DISTAL FEMUR IMPLANT LIMB SALVAGE SYSTEM KRO STANMORE IMPLANTS WORLDWIDE LTD BME6305

Patients

Seq Age Sex Outcome Treatment
1 61 YR Required Intervention