FDA Adverse Event Malfunction Summary report: N

JTS EXTENDIBLE DISTAL FEMORAL IMPLANT

MDR report key: 4373014 · Received December 24, 2014

Report

Report Number
3004105610-2014-00193
Event Type
Malfunction
Date Received
December 24, 2014
Date of Event
February 26, 2014
Report Date
February 26, 2014
Manufacturer
STANMORE IMPLANTS WORLDWIDE LTD
Product Code
KRO
PMA / PMN Number
K133152
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT UNDERWENT SUCCESSFUL REVISION SURGERY RESULTING IN THE IMPLANTATION OF A NEW DISTAL FEMUR. THE FRACTURED DISTAL FEMUR JTS WAS IN-SITU FOR 4 YEARS. THERE ARE NO ADDITIONAL DETAILS AVAILABLE WHICH DEFINITIVELY ASSIGN A ROOT CAUSE TO THE FRACTURE, HOWEVER IMPLANT FRACTURES ARE WELL RECOGNIZED, LATE POST-OPERATIVE OCCURRENCES MAY BE ASSOCIATED WITH EXCESSIVE LOADS, IMPACT OR TRAUMA IN IMPLANTED PATIENTS. THIS COMPLAINT IS BEING CLOSED, AND IS BEING TRACKED AND TRENDED.

Additional Manufacturer Narrative · 1

THE IMPLANT HAS NOT BEEN RETURNED FOR EVAL, THEREFORE, A VISUAL INSPECTION HAS NOT BEEN POSSIBLE. A REVIEW OF THE MANUFACTURING HISTORY CONFIRMS NO NON-CONFORMANCES WERE IDENTIFIED. IT IS NOT POSSIBLE TO IDENTIFY THE ROOT CAUSE OF THE REPORTED FRACTURE. REQUESTS ARE BEING MADE FOR ADD'L INFO AND A SUPPLEMENTAL REPORT WILL BE PROVIDED IF THIS INFO IS RECEIVED. PLEASE NOTE THAT THIS CUSTOM IMPLANT IS SIMILAR TO THE METS MODULAR DISTAL FEMUR (K121029).

Description of Event or Problem · 1

IT WAS REPORTED BY THE SURGEON THAT THE PATIENT IS BEING REVISED DUE TO A BROKEN STEM. THIS IS A SUPPLEMENTAL REPORT TO 3004105610-2014-00193 (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY THE SURGEON THAT THE PT IS BEING REVISED DUE TO A BROKEN STEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
850786 JTS EXTENDIBLE DISTAL FEMORAL IMPLANT LIMB SALVAGE SYSTEM KRO STANMORE IMPLANTS WORLDWIDE LTD BME15370

Patients

Seq Age Sex Outcome Treatment
1 23 YR Required Intervention