FDA Adverse Event Malfunction Summary report: N

SIMPLEX P-JAPANESE TWIN PACK

MDR report key: 4372334 · Received December 30, 2014

Report

Report Number
0002249697-2014-05037
Event Type
Malfunction
Date Received
December 30, 2014
Date of Event
December 5, 2014
Report Date
December 11, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
LOD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AN EVENT REGARDING A PIN HOLE IN AN INNER POWDER POUCH INVOLVING SIMPLEX BONE CEMENT WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD & RESULTS: DEVICE EVALUATION AND RESULTS: BASED ON THE VISUAL INSPECTION, THERE IS A PIN HOLE PRESENT NEAR THE EDGE OF THE TOP OF THE INNER POUCH. THERE ARE IMPRESSIONS ON THE POUCH ABOVE THE PIN HOLE. THESE IMPRESSIONS AND THE PIN HOLE APPEAR TO HAVE BEEN CAUSED BY A FORCEPS BEING USED TO GRIP THE INNER POUCH TO REMOVE IT IN A STERILE MANNER FROM THE OUTER POUCH IN THE OR. MEDICAL RECORDS RECEIVED AND EVALUATION: NOT PERFORMED AS THE REPORTED COMPONENT WAS NOT IMPLANTED. DEVICE HISTORY REVIEW: BMR REVIEW FOR THE REPORTED LOT WAS SATISFACTORY. COMPLAINT HISTORY REVIEW: CHR REVIEW FOR THE REPORTED LOT CONFIRMED THAT THERE WERE NO SIMILAR EVENTS REPORTED FOR THE LOT. CONCLUSIONS: BASED ON THE VISUAL INSPECTION, THE INVESTIGATION CONCLUDED THAT THE ROOT CAUSE OF THE PINHOLE TO BE AS A RESULT OF FORCEPS OR SIMILAR INSTRUMENT BEING USED TO REMOVE THE INNER POUCH FROM THE OUTER POUCH IN THE OPERATING ROOM. THE PRODUCT WAS NOT SHIPPED IN THIS CONDITION AS INDICATED BY THE VISUAL EVIDENCE.

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT PIN HOLE WAS CONFIRMED ON THE INNER PACKAGE BEFORE USED IN THE PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT PIN HOLE WAS CONFIRMED ON THE INNER PACKAGE BEFORE USED IN THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
859217 SIMPLEX P-JAPANESE TWIN PACK IMPLANT LOD STRYKER ORTHOPAEDICS-MAHWAH JFV019

Patients

Seq Age Sex Outcome Treatment
1 Other