CREATININE REAGENT
Report
- Report Number
- 9680746-2014-00003
- Event Type
- Malfunction
- Date Received
- December 29, 2014
- Date of Event
- December 3, 2014
- Report Date
- December 3, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- CGX
- PMA / PMN Number
- K934361
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CUSTOMER DID NOT PROVIDE THE PATIENT'S AGE, DATE OF BIRTH AND WEIGHT FOR THIS EVENT. THE CUSTOMER PROVIDED A REACTION MONITOR FROM THE AU5800 CLINICAL CHEMISTRY ANALYZER FOR REVIEW AND ANALYSIS INDICATES THAT THE GENERATED RESULT IS A TRUE LOW CREATININE RESULT. THERE IS NO EVIDENCE OF ANY MIXING, SAMPLING OR PHOTOMETRY ERRORS. THE CUSTOMER DID NOT REQUEST FOR SERVICE AND NO PARTS WERE REPLACED ON THE INSTRUMENT. THE EVENT ONLY AFFECTED ONE (1) PATIENT SAMPLE. IN CONCLUSION, THE CAUSE OF THE EVENT IS UNKNOWN AND THERE IS NO EVIDENCE OF A MALFUNCTION OF THE AU5800 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO INDICATION THAT CREATININE REAGENT WAS RETURNED FOR EVALUATION.
THE CUSTOMER REPORTED OBTAINING AN ERRONEOUSLY LOW CREATININE RESULT FOR ONE (1) PATIENT FROM THE AU5800 CLINICAL CHEMISTRY ANALYZER. THE CUSTOMER INDICATED THAT THE RESULT WAS NOT REPORTED OUT OF THE LABORATORY AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. THE CUSTOMER STATED THAT AN INITIAL CREATININE RESULT OF 114 UMOL/L WAS GENERATED FROM THE AU5800 CLINICAL CHEMISTRY ANALYZER. SUBSEQUENT REPEAT OF THE SAME SAMPLE ON AN ALTERNATE ANALYZER (HITACHI) GENERATED A RESULT OF 202 UMOL/L. THE CUSTOMER STATED THAT THE AU RESULT IS NOT IN ALIGNMENT WITH THE PATIENT'S CLINICAL DIAGNOSIS. THE CUSTOMER STATED THAT CALIBRATION AND QUALITY CONTROL RESULT FOR CREATININE WERE NORMAL AT THE TIME OF THE EVENT. THE CREATININE REAGENT WAS USED IN CONJUNCTION WITH THE AU5800 CLINICAL CHEMISTRY ANALYZER FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 856258 | CREATININE REAGENT | ALKALINE PICRATE, COLORIMETRY, CREATININE | CGX | BECKMAN COULTER | NA | 6241 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |