FDA Adverse Event
Injury
Summary report: N
MAQUET, MOBILE OR TABLE ALPHASTAR
MDR report key: 4365327
·
Received December 17, 2014
Report
- Report Number
- 3008355164-2014-00393
- Event Type
- Injury
- Date Received
- December 17, 2014
- Date of Event
- May 11, 2012
- Report Date
- May 11, 2012
- Manufacturer
- MAQUET GMBH,
- Product Code
- FQO
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT A MAQUET OPERATING TABLE SLOWLY LOWERED DURING A HIP SURGERY IN OPERATING ROOM NUMBER 1. THE RIGHT SIDE ADAPTER BECAME DETACHED FROM THE TABLE, WHICH ALLOWED THE LEG EXTENSION PIECE TO SLIDE SIDEWAYS OUT OF THE LEFT SIDE ADAPTER AND FALL TO THE FLOOR. THIS ALSO CAUSED THE PATIENT TO FALL. THE EVENT OCCURRED AFTER THE PATIENT WAS ANESTHETIZED AND INTUBATED, BUT PRIOR TO PREPPING AND DRAPING. THE PATIENT EXPERIENCED BRUISING AND SORENESS OF THE LOWER BACK AND ARM. THE SURGERY WAS ALSO CANCELLED DUE TO THE REPORTED PROBLEM. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 829284 | MAQUET, MOBILE OR TABLE ALPHASTAR | FQO | MAQUET GMBH, | 1132.11B2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR | Other |