FDA Adverse Event Malfunction Summary report: N

SYNCHROMED II

MDR report key: 4365277 · Received December 29, 2014

Report

Report Number
3004209178-2014-24576
Event Type
Malfunction
Date Received
December 29, 2014
Report Date
December 24, 2014
Manufacturer
MDT PUERTO RICO OPERATIONS CO
Product Code
LKK
PMA / PMN Number
P860004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: PRODUCT ID: 8731SC, SERIAL# (B)(4), IMPLANTED: (B)(6) 2010, PRODUCT TYPE: CATHETER. PRODUCT ID: 8598A, SERIAL# (B)(4), IMPLANTED: (B)(6) 2010, PRODUCT TYPE: CATHETER. PRODUCT ID: 8709, LOT# J0177999R, IMPLANTED: (B)(6) 2001, PRODUCT TYPE: CATHETER. (B)(4).

Description of Event or Problem · 1

THE PATIENT EXPERIENCED MORE SPASMS IN THE LEFT LEG COMPARED TO THE RIGHT LEG. THE PATIENT USES A CANE MOST OF THE TIME. A HEALTHCARE PROFESSIONAL (HCP) LATER REPORTED THAT THE PATIENT EXPERIENCED INCREASED SPASTICITY, THOUGH NOT CONSISTENTLY. THE PUMP HAD BEEN STALLING MORE FREQUENTLY SINCE (B)(6) 2014. THE LOGS REVEALED THE FOLLOWING: A NORMAL STALL WITH RECOVERY DUE TO MRI ON (B)(6); A NINE MINUTE STALL WITH RECOVERY ON (B)(6); A NINETEEN MINUTE STALL WITH RECOVERY ON (B)(6); A FIVE MINUTE STALL WITH RECOVERY ON (B)(6); A FIFTEEN MINUTE STALL WITH RECOVERY ON (B)(6); A FIVE MINUTE STALL WITH RECOVERY ON (B)(6); A THIRTY-FIVE MINUTE STALL WITH RECOVERY ON (B)(6); A NINE MINUTE STALL WITH RECOVERY ON (B)(6); A NINETEEN MINUTE STALL WITH RECOVERY ON (B)(6); AND A SEVEN MINUTE STALL WITH RECOVERY ON (B)(6). THEY INITIALLY THOUGHT THE STALLS WERE RELATED TO A MAGNET, BUT THEY DID NOT OCCUR AT THE CORRECT TIMES. THERE WAS REPORTEDLY NO RHYME OR REASON TO THE STALLS. THE PATIENT¿S PUMP CONTAINED LIORESAL BETWEEN (B)(6) 2014. THE PUMP CONTAINED GABLOFEN. THE LOG NOTES INDICATED THE PUMP CONTAINED LIORESAL AS EARLY AS (B)(6) 2014. THE PATIENT RECOVERED WITHOUT PERMANENT IMPAIRMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
855283 SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE LKK MDT PUERTO RICO OPERATIONS CO 8637-20

Patients

Seq Age Sex Outcome Treatment
1 00056 YR