LUCAS CPR CHEST COMPRESSION SYSTEM
Report
- Report Number
- 3015876-2014-01505
- Event Type
- Death
- Date Received
- December 17, 2014
- Date of Event
- November 5, 2014
- Report Date
- November 18, 2014
- Manufacturer
- JOLIFE AB
- Product Code
- DRM
- PMA / PMN Number
- K090422
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Location
- SW
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B)(4). BASED ON THE REPORTED INFO, THERE WAS NO INDICATION OF A DEVICE MALFUNCTION OR USE ERROR ASSOCIATED WITH THE REPORTED EVENT. THE CUSTOMER CONTINUES TO USE THE DEVICE. A CLINICAL REVIEW OF THE REPORTED ISSUE WAS PERFORMED BY JOLIFE AND PHYSIO-CONTROL WITH THE FOLLOWING CONCLUSIONS: INJURIES AT THE HEART AND GREAT VESSELS OCCUR IN ABOUT 11% OF THE CASES DURING MANUAL CPR (REFERENCE: CARDIAC ARREST, THE SCIENCE AND PRACTICE OF RESUSCITATION MEDICINE. PARADIS, HALPERIN PAGE 576). BASED ON THE LIMITED INFO PROVIDED BY THE HOSPITAL ASSOCIATE PROFESSOR REGARDING THE INJURY AT THE PERICARDIUM AND INFO FROM THE HOSPITAL FORENSIC DEPARTMENT WHICH COULD NEITHER CONFIRM OR EXCLUDE THAT THE DAMAGE WAS CAUSED BY LUCAS, IT MAY NOT BE EXCLUDED THAT THE USE OF LUCAS CONTRIBUTED TO THE PT'S OUTCOME. THE DEVICE WAS NOT RETURNED TO PHYSIO-CONTROL TO JOLIFE FOR EVAL. PHYSIO-CONTROL IS SUBMITTING THIS MEDWATCH REPORT ON BEHALF OF JOLIFE AB (MFR). (B)(4).
THE CUSTOMER REPORTED THEY RESPONDED TO A CARDIAC ARREST. THEIR LUCAS CPR CHEST COMPRESSION DEVICE (LUCAS) WAS CONNECTED TO THE PT DURING TRANSPORT IN A HELICOPTER FOR APPROX 40 MINUTES. ONCE THE PT ARRIVED AT THE HOSPITAL, THE LUCAS WAS REMOVED AND MANUAL CPR WAS CONTINUED. THE PT NEVER REGAINED CARDIAC ACTIVITY AND DID NOT SURVIVE. IT WAS LATER CONFIRMED THAT AN INJURY TO THE PT'S PERICARDIUM HAD OCCURRED. THERE WERE ALSO MULTIPLE RIB FRACTURES AND HEMOTHORAX NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 826555 | LUCAS CPR CHEST COMPRESSION SYSTEM | COMPRESSOR, CARDIAC, EXTERNAL | DRM | JOLIFE AB | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Death |