FDA Adverse Event Malfunction Summary report: N

EASYPUMP II

MDR report key: 4360514 · Received December 17, 2014

Report

Report Number
9610825-2014-00456
Event Type
Malfunction
Date Received
December 17, 2014
Report Date
November 28, 2014
Manufacturer
B. BRAUN MELSUNGEN AG
Product Code
MEB
PMA / PMN Number
K081905
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). WE RECEIVED ONE USED FILLED EASY PUMP II LT 100-50-S WITHOUT PACKAGING. WE DETECTED A CRACK AT THE FILLING PORT (LLI-CONE). IN AS-RECEIVED CONDITION THE WHITE CLAMP WAS CLOSED. THE ORIGINAL WING CAP WAS NOT RETURNED BY THE CUSTOMER. AFTER OPENING THE TOP CAP AND REMOVING THE CLOSING CONE, WE DETECTED SOLUTION (LIQUID) AND CRYSTALLIZED DRUG RESIDUES AT THE FILLING PORT (LLI-CONE). FURTHERMORE WE DETECTED NO CRYSTALLIZED DRUG RESIDUES AT THE LLA-CONE OF THE PT CONNECTOR. ADDITIONALLY THE PUMP WAS FILLED WITH NACL 0.9% UP TO THE NOMINAL VALUE (100 ML) AND A FUNCTIONAL TEST RESPECTIVELY A LEAK TEST WAS CARRIED OUT. AFTER FILLING THE PUMP AND WAITING FOR A WHILE THE PUMP WORKED (SOLUTION WAS RUNNING). LEAKAGES WERE NOT DETECTED AT THE SAMPLE. FURTHERMORE WE TESTED THE FLOW RATE OF THE RECEIVED SAMPLE. NOMINAL: 2 ML/H ACTUAL: 1, 9 ML IN 1 H; 3, 8 ML IN 2 HRS; 6, 4 ML IN 3 HRS. THE FLOW OF THE RECEIVED SAMPLE IS WITHIN OUR SPECIFICATIONS. DURING THE TEST OF THE FLOW RATE WE DETECTED NO LEAKS AT THE SAMPLE. WE HAVE INFORMED OUR MFR ACCORDINGLY. THE COMPLAINT WAS NOT JUSTIFIED AS THE SAMPLES WERE WITHIN SPECIFICATION. REVIEWED THE DEVICE HISTORY RECORD AND THERE WERE NO DEFECTS ENCOUNTERED DURING IN PROCESS AND FINAL CONTROL INVESTIGATIONS.

Description of Event or Problem · 1

AS REPORTED BY THE USER FACILITY (TRANSLATION OF USER FACILITY INFO BY (B)(4)): UNABLE TO START THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
830160 EASYPUMP II ELASTOMERIC INFUSION PUMP MEB B. BRAUN MELSUNGEN AG NA 14G08GE262

Patients

Seq Age Sex Outcome Treatment
1 UNK