FDA Adverse Event Malfunction Summary report: N

PANDA IRES INFANT WARMER

MDR report key: 4357276 · Received December 16, 2014

Report

Report Number
4357276
Event Type
Malfunction
Date Received
December 16, 2014
Date of Event
July 1, 2014
Report Date
August 13, 2014
Manufacturer
OHMEDA MEDICAL
Product Code
FMT
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MD, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

DURING A ROUTINE BIOMED PREVENTIVE MAINTENANCE, IT WAS NOTED THAT THERE WERE TWO PANDA WARMERS WITH INTERMITTENT SYSTEM FAILURES. WHEN THE SYSTEM FAILURE OCCURS, THE MESSAGE "SYSTEM FAILURE 11 AA AA AA AA" IS DISPLAYED ON THE WARMER SCREEN AND AN AUDIBLE ALARM SOUNDS. THE WARMER STOPS WARMING BUT CONTINUES DISPLAYING THE FAILURE MESSAGE WITH THE AUDIBLE ALARM. THIS CONTINUES UNTIL THE WARMER POWER IS CYCLED OFF AND BACK ON. THE WARMER MAY OR MAY NOT OPERATE PROPERLY AFTER THE POWER HAS BEEN CYCLED. AFTER WORKING WITH THE MANUFACTURER TO TROUBLESHOOT THE DEVICE, ONE OF THE WARMERS HAS FAILED AGAIN AFTER A MONTH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
825381 PANDA IRES INFANT WARMER WARMER, INFANT RADIANT FMT OHMEDA MEDICAL * *
825382 PANDA IRES INFANT WARMER WARMER, INFANT RADIANT FMT OHMEDA MEDICAL * *