PANDA IRES INFANT WARMER
MDR report key: 4357276
·
Received December 16, 2014
Report
- Report Number
- 4357276
- Event Type
- Malfunction
- Date Received
- December 16, 2014
- Date of Event
- July 1, 2014
- Report Date
- August 13, 2014
- Manufacturer
- OHMEDA MEDICAL
- Product Code
- FMT
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MD, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
DURING A ROUTINE BIOMED PREVENTIVE MAINTENANCE, IT WAS NOTED THAT THERE WERE TWO PANDA WARMERS WITH INTERMITTENT SYSTEM FAILURES. WHEN THE SYSTEM FAILURE OCCURS, THE MESSAGE "SYSTEM FAILURE 11 AA AA AA AA" IS DISPLAYED ON THE WARMER SCREEN AND AN AUDIBLE ALARM SOUNDS. THE WARMER STOPS WARMING BUT CONTINUES DISPLAYING THE FAILURE MESSAGE WITH THE AUDIBLE ALARM. THIS CONTINUES UNTIL THE WARMER POWER IS CYCLED OFF AND BACK ON. THE WARMER MAY OR MAY NOT OPERATE PROPERLY AFTER THE POWER HAS BEEN CYCLED. AFTER WORKING WITH THE MANUFACTURER TO TROUBLESHOOT THE DEVICE, ONE OF THE WARMERS HAS FAILED AGAIN AFTER A MONTH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 825381 | PANDA IRES INFANT WARMER | WARMER, INFANT RADIANT | FMT | OHMEDA MEDICAL | * | * | |
| 825382 | PANDA IRES INFANT WARMER | WARMER, INFANT RADIANT | FMT | OHMEDA MEDICAL | * | * |