ONCENTRA BRACHY TPS
Report
- Report Number
- 9611894-2014-00005
- Date Received
- December 23, 2014
- Date of Event
- November 26, 2014
- Report Date
- March 19, 2015
- Manufacturer
- NUCLETRON
- Product Code
- MUJ
- PMA / PMN Number
- K132816
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FINAL REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.
MANUFACTURER'S INVESTIGATION RESULTS: THE PROBLEM COULD NOT BE REPRODUCED IN ONCENTRA BRACHY. HOWEVER CHANNEL NUMBERING IN THE CASE EXPLORER WAS NOT UPDATED CORRECTLY. CHANNEL NUMBERING IS NOT AVAILABLE IN THE EXPORTED TREATMENT PLAN AFTER EXPORT OF THE TREATMENT PLAN TO TCS IN PROPRIETARY FORMAT. PROPRIETARY PLAN FORMATS SHOULD BE AVOIDED SINCE IT DOES NOT INCLUDE CATHETER NUMBERING INFORMATION. EXPORT IN DICOM FORMAT IS HANDLED CORRECTLY AS THIS WILL NOT AFFECT THE CHANNEL NUMBERING. ELEKTA ADVICES TO EXPORT THE TREATMENT PLAN IN DICOM FORMAT AS THE PROBLEM DOES NOT OCCUR WHEN EXPORTING THE PLAN IN DICOM FORMAT. THE LATEST RELEASE OF TCS NOW FULLY SUPPORTS DICOM PLANS, WHICH MEANS THAT THE TRANSFERRING OF PROPRIETARY PLANS IS NO LONGER NECESSARY. ALWAYS EXECUTE THE EXTRA CHECK CABLE RUN IN TCS BEFORE START OF THE TREATMENT. THIS WILL RESULT IN AN ERROR MESSAGE WHEN TRANSFER TUBES ARE NOT INSERTED INTO THE CORRECT AFTERLOADER CHANNEL. MANUFACTURER'S RECOMMENDED ACTION: ERROR MESSAGES ARE IN PLACE, INDICATING AN INCORRECT USE OF: TOTAL NUMBER OF TRANSFER TUBES. MISSING TRANSFER TUBES IN THE AFTERLOADER CHANNELS. ALWAYS EXECUTE THE EXTRA CHECK CABLE RUN IN TCS BEFORE START OF THE TREATMENT.
THE CUSTOMER REPORTED THAT OCB 4.3 CHANGED TREATMENT CHANNELS AFTER DELETING A CATHETER, LEADING TO THE MISTREATMENT OF A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 845624 | ONCENTRA BRACHY TPS | SYSTEM,PLANNING,RADIATION THERAPY TREATMENT | MUJ | NUCLETRON |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |