FDA Adverse Event Malfunction Summary report: N

BO-HIT 7050 HLS

MDR report key: 4356062 · Received December 17, 2014

Report

Report Number
8010762-2014-01371
Event Type
Malfunction
Date Received
December 17, 2014
Date of Event
November 21, 2014
Report Date
November 24, 2014
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DTZ
PMA / PMN Number
K101153
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

MAQUET CARDIOPULMONARY IS AWARE OF SIMILAR COMPLAINTS. THE DEVICES DISPLAYED A SIMILAR MALFUNCTION WHICH WERE TESTED AND EVALUATED UNDER AN OPTICAL MICROSCOPE. DELAMINATION OF SOME GAS FIBERS WAS OBSERVED WHICH ALLOWED FOR THE PRIMING SOLUTION OR BLOOD TO FLOW INSIDE THE GAP BETWEEN THE GAS FIBERS AND POLYURETHANE. GRAVITY THEN ALLOWED FOR PASSAGE TO THE GAS EXITING PATH ALONG THE HOUSING. THE MOST PROBABLE ROOT CAUSE IS THE DELAMINATION OF THE HOLLOW GAS FIBERS FROM THE POLYURETHANE POTTING AREA. A REVIEW OF THE QUAL CONTROL PROCESS CONFIRMS THAT 100% FUNCTIONAL INSPECTION FOR LEAKAGE IS PERFORMED DURING PRODUCTION. MAQUET CARDIOPULMONARY AG HAS INITIATED AN INTERNAL PROCESS ((B)(4)) TO ADDRESS THE APPROPRIATE CORRECTIVE AND PREVENTIVE ACTION. A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED WHEN NEW INFO BECOMES AVAILABLE. ADD'L INFO: THE PRODUCT MENTIONED UNDER IS A TUBING SET AND THE INCLUDED AFFECTED COMPONENT HAS THE CONTRIBUTING DESIGN FUNCTION OF THE QUADROX-ID WHICH IS REGISTERED UNDER 510 (K): K101153.

Description of Event or Problem · 1

IT WAS REPORTED THAT A LEAK WAS NOTED AT THE GAS OUTLET. BLOOD LOSS WAS APPROXIMATELY 200ML. THE DEVICE WAS NOT REPLACED. (B)(4).