BO-HIT 7050 HLS
Report
- Report Number
- 8010762-2014-01371
- Event Type
- Malfunction
- Date Received
- December 17, 2014
- Date of Event
- November 21, 2014
- Report Date
- November 24, 2014
- Manufacturer
- MAQUET CARDIOPULMONARY AG
- Product Code
- DTZ
- PMA / PMN Number
- K101153
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
MAQUET CARDIOPULMONARY IS AWARE OF SIMILAR COMPLAINTS. THE DEVICES DISPLAYED A SIMILAR MALFUNCTION WHICH WERE TESTED AND EVALUATED UNDER AN OPTICAL MICROSCOPE. DELAMINATION OF SOME GAS FIBERS WAS OBSERVED WHICH ALLOWED FOR THE PRIMING SOLUTION OR BLOOD TO FLOW INSIDE THE GAP BETWEEN THE GAS FIBERS AND POLYURETHANE. GRAVITY THEN ALLOWED FOR PASSAGE TO THE GAS EXITING PATH ALONG THE HOUSING. THE MOST PROBABLE ROOT CAUSE IS THE DELAMINATION OF THE HOLLOW GAS FIBERS FROM THE POLYURETHANE POTTING AREA. A REVIEW OF THE QUAL CONTROL PROCESS CONFIRMS THAT 100% FUNCTIONAL INSPECTION FOR LEAKAGE IS PERFORMED DURING PRODUCTION. MAQUET CARDIOPULMONARY AG HAS INITIATED AN INTERNAL PROCESS ((B)(4)) TO ADDRESS THE APPROPRIATE CORRECTIVE AND PREVENTIVE ACTION. A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED WHEN NEW INFO BECOMES AVAILABLE. ADD'L INFO: THE PRODUCT MENTIONED UNDER IS A TUBING SET AND THE INCLUDED AFFECTED COMPONENT HAS THE CONTRIBUTING DESIGN FUNCTION OF THE QUADROX-ID WHICH IS REGISTERED UNDER 510 (K): K101153.
IT WAS REPORTED THAT A LEAK WAS NOTED AT THE GAS OUTLET. BLOOD LOSS WAS APPROXIMATELY 200ML. THE DEVICE WAS NOT REPLACED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 829961 | BO-HIT 7050 HLS | HLS SET | DTZ | MAQUET CARDIOPULMONARY AG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |