PUMP IN STYLE WIC BREAST PUMP GO W/O BATTERY
Report
- Report Number
- 1419937-2014-00924
- Event Type
- Malfunction
- Date Received
- December 18, 2014
- Date of Event
- November 19, 2014
- Report Date
- December 15, 2014
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE PRODUCT WAS RECEIVED AND EVALUATED BY QE ON 01/10/2014 AND CONFIRMED A BREACH IN THE TRANSFORMER HOUSING. PRODUCT EVALUATION: EVALUATION SUMMARY: A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS BREACHED, EXPOSING INNER ELECTRICAL CIRCUITRY. A VISUAL EXAMINATION OF THE CORD REVEALED EXPOSED WIRES ON THE CORD. THE POWER SUPPLY WAS PLUGGED IN SO THE VOLTAGE ACROSS THE DC PLUG WAS MEASURED. RESISTANCE ACROSS THE DC PLUG WAS MEASURED 0.760M OHMS, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. THE RESISTANCE ACROSS AC BLADES MEASURED AT 54.9 OHMS, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW. THIS ISSUE WITH A DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING., HANDLING AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STAKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING.
THE CUSTOMER REPORTED A DAMAGED FACEPLATE. UPON RECEIPT OF THE PRODUCT AND QE EVAL A BREACH IN THE TRANSFORMER HOUSING WAS OBSERVED, WHICH IS A SAFETY RISK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 834640 | PUMP IN STYLE WIC BREAST PUMP GO W/O BATTERY | HGX | MEDELA, INC. | 9207010/57018W | REV L-2011 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |