FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE WIC BREAST PUMP GO W/O BATTERY

MDR report key: 4355703 · Received December 18, 2014

Report

Report Number
1419937-2014-00924
Event Type
Malfunction
Date Received
December 18, 2014
Date of Event
November 19, 2014
Report Date
December 15, 2014
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RECEIVED AND EVALUATED BY QE ON 01/10/2014 AND CONFIRMED A BREACH IN THE TRANSFORMER HOUSING. PRODUCT EVALUATION: EVALUATION SUMMARY: A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS BREACHED, EXPOSING INNER ELECTRICAL CIRCUITRY. A VISUAL EXAMINATION OF THE CORD REVEALED EXPOSED WIRES ON THE CORD. THE POWER SUPPLY WAS PLUGGED IN SO THE VOLTAGE ACROSS THE DC PLUG WAS MEASURED. RESISTANCE ACROSS THE DC PLUG WAS MEASURED 0.760M OHMS, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. THE RESISTANCE ACROSS AC BLADES MEASURED AT 54.9 OHMS, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW. THIS ISSUE WITH A DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING., HANDLING AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STAKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A DAMAGED FACEPLATE. UPON RECEIPT OF THE PRODUCT AND QE EVAL A BREACH IN THE TRANSFORMER HOUSING WAS OBSERVED, WHICH IS A SAFETY RISK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
834640 PUMP IN STYLE WIC BREAST PUMP GO W/O BATTERY HGX MEDELA, INC. 9207010/57018W REV L-2011

Patients

Seq Age Sex Outcome Treatment
1