FDA Adverse Event Malfunction Summary report: N

STANDING AND RAISING ASSIST-S.A.R.A. LIFT WITH BACK BELT

MDR report key: 4352 · Received June 24, 1992

Report

Report Number
4352
Event Type
Malfunction
Date Received
June 24, 1992
Date of Event
January 17, 1992
Report Date
January 20, 1992
Manufacturer
ARJO-CENTURY, INC.
Product Code
INE
Product Problem
Yes
Report Source
User Facility report

Narratives

Description of Event or Problem · 1

C.N.A. WAS LOWERING RESIDENT WITH S.A.R.A. LIFT AND CRANK (HANDLE) SWUNG AND HIT C.N.A.'S ELBOW. THIS TRANSFERRING DEVICE HAS A LARGE HANDLE WITH THREE HAND GRIPS, ONE WAS MISSING AND HAD BEEN ORDERED PRIOR TO INCIDENT, TO DRIVE THE SLING TYPE LIFTING MECHANISM. PROPER INSTRUCTIONS AND PROCEDURE TO TURN HANDLE SLOWLY WERE NOT IMPLEMENTED FACILITATING HANDLE SWINGING AND INJURY UPON EVALUATION. PHYSICIAN ORDERED NO LIFTING FOR 48 HOURS DUE TO LEFT (L) ELBOW BRUISING AND SWELLING FOR C.N.A. UPON EXAMINATIONDEVICE NOT LABELED FOR SINGLE USE. PATIENT MEDICAL STATUS PRIOR TO EVENT: SATISFACTORY CONDITION. THERE WAS NOT MULTIPLE PATIENT INVOLVEMENT.DEVICE SERVICED IN ACCORDANCE WITH SERVICE SCHEDULE. DATE LAST SERVICED: 01-MAY-91. SERVICE PROVIDED BY: FACTORY TRAINED/AUTHORIZED/OWNED SERVICE ORGANIZATION. SERVICE RECORDS AVAILABLE.NO IMMINENT HAZARD TO PUBLIC HEALTH CLAIMED. DEVICE USED AS LABELED/INTENDED.DEVICE WAS EVALUATED AFTER THE EVENT. METHOD OF EVALUATION: ACTUAL DEVICE INVOLVED IN INCIDENT WAS EVALUATED, MECHANICAL TESTS PERFORMED, PERFORMANCE TESTS PERFORMED. RESULTS OF EVALUATION: FAILURE TO FOLLOW INSTRUCTIONS, UNANTICIPATED SHORT TERM COMPLICATION OF PROCEDURE. CONCLUSION: DEVICE WAS OUT OF SPEC IN A MANNER THAT RELATES TO EVENT, USER ERROR CONTRIBUTED TO EVENT. CERTAINTY OF DEVICE AS CAUSE OF OR CONTRIBUTOR TO EVENT: YES. CORRECTIVE ACTIONS: DEVICE REPAIRED AND PUT BACK IN SERVICE, USER EDUCATION PROVIDED. INVALID DATA - ON DEVICE DESTROYED/DISPOSED OF STATUS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STANDING AND RAISING ASSIST-S.A.R.A. LIFT WITH BACK BELT INE ARJO-CENTURY, INC. KB1003

Patients

Seq Age Sex Outcome Treatment
1 17 YR Other